文献速递胸部技术与安全US

人胸壁超声衰减的实验测量及肺超声机械指数的评估

Experimental Measurements of Ultrasound Attenuation in Human Chest Wall and Assessment of the Mechanical Index for Lung Ultrasound.
PubMed 文献 · Ultrasound Med Biol · 2020-03-24

研究测量了15例人尸体胸壁肋间组织在1.6–5 MHz下的超声衰减系数,样本中位厚度3.8 cm(范围2.3–5.5 cm),频率归一化衰减系数约1.44 dB/cm/MHz,样本平均线性衰减系数为1.43 ± 0.32 dB/cm/MHz;该衰减高于小鼠和大鼠胸壁的1.1–1.3 dB/cm/MHz,也远大于机械指数采用的0.3 dB/cm/MHz。用0.3 dB/cm/MHz降额盐水值计算的机械指数比经组织实测波形计算值最多高1.2 MPa/MHz1/2,作者认为机械指数高估了肺超声的暴露,可能不是肺超声适用的剂量学指标。

要点

  • 人胸壁肋间组织频率归一化衰减系数约1.44 dB/cm/MHz,样本平均线性衰减系数1.43 ± 0.32 dB/cm/MHz。
  • 该衰减高于小鼠和大鼠胸壁的1.1–1.3 dB/cm/MHz,远大于机械指数采用的0.3 dB/cm/MHz。
  • 机械指数高估肺超声暴露,可能不是肺超声合适的剂量学指标。

核心结论

  • 人胸壁超声衰减高于既往哺乳动物胸壁测量值,且远大于机械指数采用的0.3 dB/cm/MHz。
  • 机械指数高估了肺超声的暴露,可能不是肺超声合适的剂量学指标。

研究设计与样本

  • 实验测量研究,使用15例人尸体胸壁样本,测量经肋间组织相对于盐水的超声透射以确定各样本衰减系数。
  • 使用GE Vivid 7诊断超声仪(GE Vingmed,Horten,Norway)和3 S与5 S相控阵探头,中心频率1.6至5 MHz。
  • 胸壁样本厚度2.3–5.5 cm,中位厚度3.8 cm。

诊断标准与关键数值

  • 基于所有衰减测量值的线性最佳拟合,频率归一化衰减系数约为1.44 dB/cm/MHz。
  • 样本平均线性衰减系数为1.43 ± 0.32(均值 ± 标准差)dB/cm/MHz。
  • 既往哺乳动物胸壁样本测量值为小鼠和大鼠1.1–1.3 dB/cm/MHz,机械指数采用0.3 dB/cm/MHz。
  • 用0.3 dB/cm/MHz降额的盐水值计算的机械指数比经组织实测超声波形计算值最多高1.2 MPa/MHz1/2。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Knowledge of the acoustic attenuation characteristics of the chest wall is necessary to estimate the acoustic exposure at the pleural surface during lung ultrasound and is useful in the prediction of bio-effects (e.g., pulmonary capillary hemorrhage) and the development of safe, effective lung imaging. Currently, this property is not well characterized in humans. The aim of this work was to characterize ultrasonic attenuation in human chest wall such that the ultrasound exposures of the lung can be estimated for clinically relevant conditions. In this study, we experimentally measured ultrasound transmitted through the intercostal tissue of 15 human cadaver chest wall samples relative to ultrasound transmitted through saline to determine attenuation coefficients for each sample. A GE Vivid 7 diagnostic ultrasound machine (GE Vingmed, Horten, Norway) and 3 S and 5 S phased array probes were used at center frequencies from 1.6 to 5 MHz. The chest wall samples varied in thickness from 2.3-5.5 cm with a median thickness of 3.8 cm. The frequency-normalized attenuation coefficient was approximately 1.44 dB/cm/MHz based on a linear best fit through all attenuation measurements. Attenuation characteristics varied appreciably between samples, and the sample-averaged linear attenuation coefficient was 1.43 ± 0.32 (mean ± standard deviation) dB/cm/MHz. This attenuation is higher than that previously measured in mammalian chest wall samples (1.1-1.3 dB/cm/MHz for mice and rats) and is much greater than that used by the mechanical index (0.3 dB/cm/MHz). Mechanical index values calculated using saline values de-rated by 0.3 dB/cm/MHz were up to 1.2 MPa/MHz1/2 greater than those calculated using the measured through-tissue ultrasound waves. We conclude that the mechanical index overestimates exposures for lung ultrasound and thus may not be an appropriate dosimetry metric for pulmonary ultrasound.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Patterson B, Miller DL. Experimental Measurements of Ultrasound Attenuation in Human Chest Wall and Assessment of the Mechanical Index for Lung Ultrasound. Ultrasound Med Biol. 2020. PMID: 32217030. doi:10.1016/j.ultrasmedbio.2020.01.031