MRI射频能量与非条件性植入心脏装置功能的关系:一项前瞻性评估

The Relationship between MRI Radiofrequency Energy and Function of Nonconditional Implanted Cardiac Devices: A Prospective Evaluation.
PubMed 文献 · Radiology · 2020-03-03

这项前瞻性观察队列研究纳入1464例受试者的2028次MRI检查,评估全身平均SAR峰值、dB/dt、最大比能量剂量、成像区域及装置特征与装置功能变化的关系。结果未发现射频能量沉积、dB/dt或扫描时长与装置参数变化相关;胸部MRI与检查后即刻电池电压降低相关(β=-0.008 V,P<.001)。作者指出MRI前后装置 interrogation 仍属必需。

要点

  • 射频能量沉积、dB/dt和扫描时长与装置参数变化无关联证据。
  • 胸部MRI与检查后即刻电池电压降低相关(β=-0.008 V,P<.001)。
  • 右心室导线长度与检查后即刻右心室感知降低(β=-0.012 mV,P=.05)和起搏阈值降低(β=-0.002 V,P<.01)相关。

核心结论

  • MRI中表征患者射频能量暴露的参数与检查后即刻装置及导线参数变化之间无关联证据。
  • 由于存在装置重置及导线或发生器参数变化的可能,MRI前后装置 interrogation 仍属必需。

研究设计与样本

  • 前瞻性观察性队列研究,2003年10月16日至2015年1月22日进行,注册号NCT01130896。
  • 共纳入1464例受试者,完成2028次MRI检查,涉及2755根装置导线;平均年龄67岁±15(标准差),男性930例(64%)。
  • 纳入标准为任何被转诊行MRI的植入心脏装置个体;使用无SAR限制的临床MRI方案。
  • 排除标准为新植入导线、废弃或心外膜导线、依赖起搏器且植入式心律转复除颤器无异步起搏功能者。
  • 对每个MRI脉冲序列收集计算的全身体SAR、dB/dt和扫描时长;每次MRI检查前及检查后即刻(10分钟内)评估心房和心室感知、导线阻抗及起搏阈值。

临床应用与阅片要点

  • 作者指出,由于存在装置重置及导线或发生器参数变化的可能,MRI前后装置 interrogation 仍属必需。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background The risks associated with MRI in individuals who have implanted cardiac devices are thought to arise from the interaction between the implanted device and static, gradient, and radiofrequency magnetic fields. Purpose To determine the relationship between the peak whole-body averaged specific absorption rate (SAR) and change in magnetic field per unit time (dB/dt), maximum specific energy dose, imaging region, and implanted cardiac device characteristics and their function in patients undergoing MRI. Materials and Methods This prospective observational cohort study was conducted from October 16, 2003, to January 22, 2015 (https://ClinicalTrials.gov, NCT01130896). Any individual with an implanted cardiac device who was referred for MRI was included. Clinical MRI protocols without SAR restriction were used. Exclusion criteria were newly implanted leads, abandoned or epicardial leads, and dependence on a pacemaker with an implantable cardioverter defibrillator without asynchronous pacing capability. For each MRI pulse sequence, the calculated whole-body values for SAR, dB/dt, and scan duration were collected. Atrial and ventricular sensing, lead impedance, and capture threshold were evaluated before and immediately after (within 10 minutes) completion of each MRI examination. Generalized estimating equations with Gaussian family, identity link, and an exchangeable working correlation matrix were used for statistical analysis. Results A total of 2028 MRI examinations were performed in 1464 study participants with 2755 device leads (mean age, 67 years ± 15 [standard deviation]; 930 men [64%]). There was no evidence of an association between radiofrequency energy deposition, dB/dt, or scan duration and changes in device parameters. Thoracic MRI was associated with decreased battery voltage immediately after MRI (β = -0.008 V, P < .001). Additionally, right ventricular (RV) lead length was associated with decreased RV sensing (β = -0.012 mV, P = .05) and reduced RV capture threshold (β = -0.002 V, P < .01) immediately after MRI. Conclusion There was no evidence of an association between MRI parameters that characterize patient exposure to radiofrequency energy and changes in device and lead parameters immediately after MRI. Nevertheless, device interrogation before and after MRI remains mandatory due to the potential for device reset and changes in lead or generator parameters. © RSNA, 2020 See also the editorial by Shellock in this issue.