CT有效剂量与卷积核对不同人工智能软件系统检测肺结节的影响:体模研究
研究用Lungman体模模拟不同大小和密度的肺结节,在4个有效剂量水平和3种卷积核下重建CT图像,用4款人工智能软件系统(A、B、C、D)检测结节数目和体积。系统B中位敏感度最高(100%),系统D相对体积误差最小(13.12%);有效剂量低于0.5 mSv时系统D敏感度下降、系统A和B假阳性和假阴性显著增加(P < 0.05)。
要点
- 系统B中位敏感度最高(100%),系统B和D中位假阳性均为1,低于系统A(11.5)和系统C(5)。
- 系统D相对体积误差最小(13.12%);卷积核由B80f改为B30f时系统A、B、C相对体积误差增加(P < 0.05)。
- 有效剂量低于0.5 mSv时系统D敏感度下降,系统A和B假阳性与假阴性显著增加(P < 0.05)。
诊断标准与关键数值
- 系统B中位敏感度最高(100%)。
- 系统B和D的中位假阳性均为1,低于系统A(11.5)和系统C(5)。
- 系统D相对体积误差最小(13.12%)。
- 有效剂量<0.5 mSv时系统D敏感度下降,系统A和B假阳性与假阴性显著增加(P < 0.05)。
- 卷积核由B80f改为B30f时系统A假阳性减少、系统C假阳性增加,系统A、B、C相对体积误差增加(P < 0.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To investigate the effective dose (E) and convolution kernel's effects on the detection of pulmonary nodules in different artificial intelligence (AI) software systems.
METHODS: Simulated nodules of various sizes and densities in the Lungman phantom were CT scanned at different levels of E (3 - 5, 1 - 3, 0.5 - 1, and <0.5 mSv) and were reconstructed with different kernels (B30f, B60f, and B80f). The number of nodules and corresponding volumes in different images were detected by four AI software systems (A, B, C, and D). Sensitivity, false positives (FPs), false negatives (FNs), and relative volume error (RVE) were calculated and compared to the aspects of the E and convolution kernel.
RESULTS: System B had the highest median sensitivity (100 %). The median FPs of systems B (1) and D (1) was lower than A (11.5) and C (5). System D had the smallest RVE (13.12 %). When the E was <0.5 mSv, system D's sensitivity decreased, while the FPs and FNs of systems A and B increased significantly (P < 0.05). When the kernel was changed from B80f to B30f, the FPs of system A decreased, while that of system C increased, and the RVE of systems A, B, and C increased (P < 0.05).
CONCLUSION: AI software systems B and D have high detection efficiency under normal or low dose conditions and show better stability. However, the detection efficiency of systems A and C would be affected by the E or convolution kernel, but the E would not affect the volume measurement of four systems.