数字PET/CT用于儿童18F-FDG脑PET重建与采集时间的临床评估
本研究回顾性纳入25例0.5~16岁癫痫患儿共400组脑PET/CT图像,比较传统迭代重建与不同β值BSREM在不同采集时间(45、90、180、300秒)下的图像质量。BSREM在各计数水平均改善质量与临床评分,低计数条件下获益最大;在合适惩罚因子(β=350~500)下,180秒与300秒采集的最高等级重建之间无统计学显著差异,提示计数密度降低40%时仍可保持诊断质量。
要点
- BSREM重建在各计数水平均改善图像质量与临床评分,低计数条件下获益最大。
- 300秒采集时总体质量评分最高见于β=500,空间分辨率评分最高见于β=200。
- 采用β=350~500时,180秒与300秒采集的最高等级重建之间无统计学显著差异。
关键影像表现/征象
- BSREM重建在所有计数水平均改善质量评分与临床评分,低计数条件下获益最大。
- 300秒采集时总体、噪声、伪影及皮层、基底节和丘脑图像质量评分最高见于β=500。
- 300秒采集时空间分辨率评分最高见于β=200。
诊断标准与关键数值
- 采用5分制评分技术质量(噪声、空间分辨率、伪影)、临床质量(皮层、基底节和丘脑图像质量)及总体诊断满意度。
- 采用β=350~500的BSREM时,180秒与300秒采集的最高等级重建之间无统计学显著差异。
- BSREM(β=350~500)在计数密度降低40%时图像质量仍具诊断性。
临床应用与阅片要点
- 儿童18F-FDG脑PET阅片与报告可关注BSREM重建参数β值对噪声、空间分辨率和伪影评分的影响。
- 在低计数条件下BSREM对图像质量的改善最明显,提示可据此评估降低计数密度后的图像可接受性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: 18F-2-fluoro-2-deoxyglucose (FDG) positron emission tomography (PET) plays an important role in the diagnosis, evaluation and treatment of childhood epilepsy. The selection of appropriate acquisition and reconstruction parameters, however, can be challenging with the introduction of advanced hardware and software functionalities.
OBJECTIVE: To quantify the diagnostic performance of a block-sequential regularized expectation maximization (BSREM) tool and reduced effective counts in brain PET/CT for pediatric epilepsy patients on a digital silicon photomultiplier system.
MATERIALS AND METHODS: We included 400 sets of brain PET/CT images from 25 pediatric patients (0.5-16 years old) in this retrospective study. Patient images were reconstructed with conventional iterative techniques or BSREM with varied penalization factor (β), at varied acquisition time (45 s, 90 s, 180 s, 300 s) to simulate reduced count density. Two pediatric nuclear medicine physicians reviewed images in random order - blinded to patient, reconstruction method and imaging time - and scored technical quality (noise, spatial resolution, artifacts), clinical quality (image quality of the cortex, basal ganglia and thalamus) and overall diagnostic satisfaction on a 5-point scale.
RESULTS: Reconstruction with BSREM improved quality and clinical scores across all count levels, with the greatest benefits in low-count conditions. Image quality scores were greatest at 300-s acquisition times with β=500 (overall; noise; artifacts; image quality of the cortex, basal ganglia and thalamus) or β=200 (spatial resolution). No statistically significant difference in the highest graded reconstruction was observed between imaging at 180 s and 300 s with an appropriately implemented penalization factor (β=350-500), indicating that a reduction in dose or acquisition time is feasible without reduction in diagnostic satisfaction.
CONCLUSION: Clinical evaluation of pediatric 18F-FDG brain PET image quality was shown to be diagnostic at reductions of count density by 40% using BSREM with a penalization factor of β=350-500. This can be accomplished while maintaining confidence of achieving a diagnostic-quality image.