利用机器学习加速的迭代重建改善CT低对比度可检测性与噪声纹理模式:体模研究

Improving Low-contrast Detectability and Noise Texture Pattern for Computed Tomography Using Iterative Reconstruction Accelerated with Machine Learning Method: A Phantom Study.
PubMed 文献 · Acad Radiol · 2020-01-07

采用CCT189 MITA CT IQ低对比度体模,在64排CT上以120 kVp、45–600 mA行螺旋扫描,比较FBP与五种迭代渐进重建(IPV)算法在0.625 mm与5.0 mm层厚下的噪声功率谱及低对比度可检测性。IPV图像归一化NPS曲线与FBP相似;0.625 mm层厚、等辐射剂量下IPV低对比度可检测性较FBP提高1.19–2.15倍,并可在维持相近低对比度可检测性时实现>50%辐射剂量降低。

要点

  • IPV图像噪声功率谱及归一化噪声功率谱曲线与FBP相似
  • 0.625 mm层厚等剂量下IPV低对比度可检测性较FBP提高1.19–2.15倍
  • IPV图像可在维持相近低对比度可检测性时实现>50%辐射剂量降低

核心结论

  • IPV图像的低对比度可检测性得到改善,其归一化NPS与FBP图像相似。
  • IPV图像在维持相近低对比度可检测性的同时实现了>50%的辐射剂量降低。

研究设计与样本

  • 体模研究,使用CCT189 MITA CT IQ低对比度体模,64排CT行螺旋扫描,管电压120 kVp、管电流45–600 mA。
  • 图像以0.625 mm和5.0 mm层厚重重建,重建算法为FBP及五种迭代渐进重建(IPV)算法。
  • 计算噪声功率谱(NPS)与归一化NPS,并采用通道化Hotelling观察者模型评估体模低对比度模块的低对比度可检测性。

关键影像表现/征象

  • IPV图像的NPS与归一化NPS曲线与FBP图像相似。
  • 在0.625 mm层厚、等辐射剂量水平下,IPV图像低对比度可检测性平均较FBP图像提高1.19–2.15倍,并伴有相应噪声降低。
  • 在5.0–8.0 HU等噪声水平下,IPVstd2至IPVstr2图像的低对比度可检测性与FBP图像几乎相同或更优;IPVstr4图像的可检测性低于FBP图像(p = 0.02)。

诊断标准与关键数值

  • 低对比度可检测性以通道化Hotelling观察者模型评估,评价对象为体模低对比度模块。
  • 0.625 mm层厚、等辐射剂量下IPV图像低对比度可检测性较FBP提高1.19–2.15倍。
  • 5.0–8.0 HU等噪声水平下IPVstd2至IPVstr2可检测性与FBP几乎相同或更优,IPVstr4低于FBP(p = 0.02)。

临床应用与阅片要点

  • 作者结论认为IPV图像在归一化NPS与FBP相似的前提下改善了低对比度可检测性,并可在维持相近低对比度可检测性时实现>50%辐射剂量降低。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To evaluate the performance of iterative reconstruction (IR) and filtered back projection (FBP) images in terms of low-contrast detectability at different radiation doses, IR levels, and slice thickness using the mathematical model observer with a focus on low-contrast detectability.

MATERIALS AND METHODS: The CCT189 MITA CT IQ Low-Contrast Phantom was used and helical scans were performed using a 64-detector CT scanner. Tube voltage was set at 120 kVp and tube current was adjusted from 45 to 600 mA. Images were reconstructed at slice thicknesses of 0.625 and 5.0 mm with FBP and five types of iterative progressive reconstruction with visual modeling (IPV) algorithms. The noise power spectrum (NPS) and normalized NPS were calculated. To evaluate low-contrast detectability, the model observer with the channelized Hotelling observer model was applied using low-contrast modules in the phantom.

RESULTS: The NPS and normalized NPS for IPV images had similar curves as that for FBP images. At a slice thickness of 0.625 mm and equivalent radiation dose level, the mean improvement of low-contrast detectability for IPV images was 1.19-2.15-fold greater than FBP images with corresponding noise reduction levels. At equivalent noise levels of 5.0-8.0 HU, low-contrast detectability of the IPVstd2 to IPVstr2 images as almost the same or better than that of the FBP images. However, the detectability of the IPVstr4 image was lower than that of the FBP image (p = 0.02).

CONCLUSION: Low-contrast detectability of the IPV images was improved with a similar normalized NPS as with FBP images. Furthermore, a radiation reduction of >50% was achieved for the IPV images, while maintaining similar low-contrast detectability.

来源:Acad Radiol,查看原文。版权归原作者所有。