使用钨功能纸与器官基管电流调制降低头部CT眼晶状体辐射剂量

Radiation dose reduction to the eye lens in head CT using tungsten functional paper and organ-based tube current modulation.
PubMed 文献 · Eur J Radiol · 2020-01-07

在头部体模上评估钨功能纸(TFP)屏蔽与器官基管电流调制(TCM)单独或联合降低眼晶状体剂量的效果。与原始扫描相比,TFP屏蔽、TCM及两者联合的剂量降低率分别为17.8%、13.6%和27.7%。联合方案在TFP与皮肤间保留空气间隙时维持图像质量。

要点

  • TFP屏蔽联合TCM可使眼晶状体剂量较原始扫描降低27.7%。
  • 前部图像噪声在联合方案下为5.0±0.2,原始扫描为4.1±0.2。
  • 增大TFP与眼距离可显著降低联合方案的CT值(p<.05)。

诊断标准与关键数值

  • 与原始扫描相比,TFP屏蔽、TCM及TFP屏蔽联合TCM的剂量降低率分别为17.8%、13.6%和27.7%。
  • 前部图像噪声(均值±标准差)在原始扫描、TFP屏蔽、TCM及联合方案分别为4.1±0.2、4.6±0.2、4.4±0.3和5.0±0.2。
  • 前部CT值(均值±标准差)在原始扫描、TFP屏蔽、TCM及联合方案分别为19.3±0.8、23.8±0.8、19.6±0.8和24.1±0.8。

临床应用与阅片要点

  • 作者采用TFP屏蔽联合TCM降低头部CT眼晶状体剂量,并强调TFP与皮肤表面间保留空气间隙以维持图像质量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: We investigated whether a tungsten functional paper (TFP) shield and/or organ-based tube current modulation (TCM) can reduce the dose to the eye lens.

MATERIALS AND METHODS: All scans were performed using our routine head examination protocol (spiral acquisition, 120 kVp, noise Index 3.5) with an anthropomorphic head phantom. The dose reduction rate was measured by the following methods with a scintillation fiber optic dosimeter: (a) without any dose reduction techniques (Original scan), (b) TFP shield, (c) TCM, and (d) TFP shield plus TCM. Image noise and CT number were obtained and compared between the three groups. In addition, image noise in method (d) was measured with varying distances between the TFP shield and eye lens.

RESULTS: The reduction rates using TFP shield, TCM, and TFP shield plus TCM compared with those for the Original scan were 17.8 %, 13.6 %, and 27.7 %, respectively. Image noise (mean ± standard deviation) in the anterior region for the Original scan, TFP shield, TCM, and TFP shield plus TCM were 4.1 ± 0.2, 4.6 ± 0.2, 4.4 ± 0.3, and 5.0 ± 0.2, while the CT numbers were 19.3 ± 0.8, 23.8 ± 0.8, 19.6 ± 0.8, and 24.1 ± 0.8, respectively. Increasing the distance between the TFP and the eye significantly decreased the CT number when using TFP shield plus TCM (p < .05).

CONCLUSION: TFP shield plus TCM reduced the dose to the eye lens in head CT while maintaining image quality with an air gap between the TFP and skin surface.