乳腺剪切波弹性成像:质量图在诊断及预测乳腺癌生物学特征中的附加价值
本研究纳入368例病理证实的乳腺病灶(均表现为剪切波弹性成像质量图中的低质量区域),比较有无疑量图引导下测量剪切波速度的诊断效能。质量图引导下平均剪切波速度的敏感度(89%)和AUC(0.932)优于无引导(敏感度84.2%,AUC 0.912),且质量图引导下平均剪切波速度与多项侵袭性癌生物学特征相关。
要点
- 在剪切波弹性成像质量图提示低质量区域时,按质量图引导避开低质量区测量可提高平均剪切波速度对乳腺病灶的诊断敏感度和AUC。
- 质量图引导下平均剪切波速度升高与脉管侵犯、腋窝淋巴结转移、雌激素受体阴性、孕激素受体阴性、人表皮生长因子受体阳性及侵袭性分子亚型相关。
- 无质量图引导时仅脉管侵犯与平均剪切波速度升高相关。
诊断标准与关键数值
- 质量图引导下平均剪切波速度诊断乳腺病灶的敏感度为89%,AUC为0.932,优于无质量图引导的敏感度84.2%、AUC 0.912(均p<0.05)。
- 质量图引导下标准差与无质量图引导下标准差之间敏感度和特异度均无显著差异(均p=1.000)。
- 质量图引导下平均剪切波速度为3.6±1.39 m/s,无质量图引导下为3.29±1.22 m/s(p<0.001)。
- 质量图引导下标准差为1.02±0.84,无质量图引导下为1.46±1.06(p<0.001)。
临床应用与阅片要点
- 当剪切波弹性成像质量图显示病灶为低质量区域时,可考虑按质量图引导避开低质量区放置感兴趣区测量剪切波速度。
- 质量图引导下平均剪切波速度可能有助于提示侵袭性乳腺癌的脉管侵犯、腋窝淋巴结转移、雌激素受体阴性、孕激素受体阴性、人表皮生长因子受体阳性及侵袭性分子亚型。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To determine the added value of a shear-wave elastography (SWE) quality map (QM) in the diagnosis of breast lesions and in predicting the biological characteristics of invasive breast cancer.
MATERIALS AND METHODS: Between January 2016 and February 2019, this study included 368 women with 368 pathologically proven breast lesions, which appeared as poor-quality regions in the QM of SWE. To measure shear-wave velocity (SWV), seven regions of interest were placed in each lesion with and without QM guidance. Under QM guidance, poor-quality areas were avoided. Diagnostic performance was calculated for mean SWV (SWVmean), max SWV (SWVmax), and standard deviation (SD) with QM guidance (SWVmean + QM, SWVmax + QM, and SD + QM, respectively) and without QM guidance (SWVmean - QM, SWVmax - QM, and SD - QM, respectively). For invasive cancers, the relationship between SWV findings and biological characteristics was investigated with and without QM guidance.
RESULTS: Of the 368 women (mean age, 47 years; SD, 10.8 years) enrolled, 159 had benign breast lesions and 209 had malignant breast lesions. SWVmean + QM (3.6 ± 1.39 m/s) and SD + QM (1.02 ± 0.84) were significantly different from SWVmean - QM (3.29 ± 1.22 m/s) and SD - QM (1.46 ± 1.06), respectively (all p < 0.001). For differential diagnosis of breast lesions, the sensitivity and areas under the receiver operating characteristic curve (AUC) of SWVmean + QM (sensitivity: 89%; AUC: 0.932) were better than those of SWVmean - QM (sensitivity, 84.2%; AUC, 0.912) (all p < 0.05). There was no significant difference in sensitivity and specificity between SD + QM and SD - QM (all p = 1.000). Among the biological characteristics of invasive cancers, lymphovascular involvement, axillary lymph node metastasis, negative estrogen receptor status, negative progesterone receptor status, positive human epidermal growth factor receptor status, and aggressive molecular subtypes showed higher SWVmean + QM (all p < 0.05), while only lymphovascular involvement showed higher SWVmean - QM (p = 0.036).
CONCLUSION: The use of QM in SWE might improve the diagnostic performance for breast lesions and facilitate prediction of the biological characteristics of invasive breast cancers.
引用
- Zheng X, Huang Y, Liu Y, et al. Shear-Wave Elastography of the Breast: Added Value of a Quality Map in Diagnosis and Prediction of the Biological Characteristics of Breast Cancer. Korean J Radiol. 2020. PMID: 31997592. doi:10.3348/kjr.2019.0453
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。