自动乳腺超声系统用于乳腺癌评估:小乳房女性中双视图扫描技术的诊断性能
研究纳入136例小乳房(胸罩罩杯A)无症状女性共272只乳房,每侧乳房分别采用双视图扫描技术(2-VST)或传统三视图扫描技术(3-VST)进行自动乳腺超声系统(ABUS)检查,并与手持超声系统(HHUS)对照。2-VST与3-VST的乳房覆盖满意率分别为94.1%和91.9%(p=0.318),敏感度均为100%(8/8 vs 10/10),特异度分别为97.7%(125/128)和95.2%(120/126),差异无统计学意义(p>0.05)。结论认为在小乳房女性中2-VST的扫描覆盖和诊断性能与3-VST相当。
要点
- 小乳房女性中ABUS双视图扫描的乳房覆盖满意率为94.1%,与三视图扫描的91.9%无显著差异(p=0.318)。
- 双视图与三视图扫描的敏感度均为100%(8/8 vs 10/10),特异度分别为97.7%(125/128)和95.2%(120/126)。
- HHUS检出的所有病灶在ABUS图像上均可见,与扫描技术无关。
核心结论
- 在小乳房女性中,ABUS双视图扫描技术(2-VST)的扫描覆盖和诊断性能与常规三视图扫描技术(3-VST)相当。
研究设计与样本
- 研究于2016年3月至2017年5月纳入136例无症状小乳房(胸罩罩杯A)适合2-VST的女性,共272只乳房。
- 每例患者双侧乳房均行ABUS和手持超声系统(HHUS)全乳超声检查;ABUS采图时一侧乳房用2-VST、另一侧用3-VST扫描。
- 评估每侧乳房的乳房覆盖情况及HHUS检出病灶在ABUS上的可见性,并比较2-VST与3-VST的敏感度和特异度。
关键影像表现/征象
- 2-VST检出8例乳腺癌,3-VST检出10例乳腺癌。
- HHUS检出的所有病灶在ABUS图像上均可见,与扫描技术无关。
诊断标准与关键数值
- 乳房覆盖满意率:2-VST为94.1%,3-VST为91.9%(p=0.318)。
- 敏感度:2-VST为100%(8/8),3-VST为100%(10/10)。
- 特异度:2-VST为97.7%(125/128),3-VST为95.2%(120/126);两者差异无统计学意义(p>0.05)。
临床应用与阅片要点
- 在小乳房(胸罩罩杯A)无症状女性中,ABUS采用2-VST可获得与3-VST相当的扫描覆盖和诊断性能。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To comparatively evaluate the scan coverage and diagnostic performance of the two-view scan technique (2-VST) of the automated breast ultrasound system (ABUS) versus the conventional three-view scan technique (3-VST) in women with small breasts.
MATERIALS AND METHODS: Between March 2016 and May 2017, 136 asymptomatic women with small breasts (bra cup size A) suitable for 2-VST were enrolled. Subsequently, 272 breasts were subjected to bilateral whole-breast ultrasound examinations using ABUS and the hand-held ultrasound system (HHUS). During ABUS image acquisition, one breast was scanned with 2-VST, while the other breast was scanned with 3-VST. In each breast, the breast coverage and visibility of the HHUS detected lesions on ABUS were assessed. The sensitivity and specificity of ABUS were compared between 2-VST and 3-VST.
RESULTS: Among 136 breasts, eight cases of breast cancer were detected by 2-VST, and 10 cases of breast cancer were detected by 3-VST. The breast coverage was satisfactory in 94.1% and 91.9% of cases under 2-VST and 3-VST, respectively (p = 0.318). All HHUS-detected lesions were visible on the ABUS images regardless of the scan technique. The sensitivities and specificities were similar between 2-VST and 3-VST (100% [8/8] vs. 100% [10/10], and 97.7% [125/128] vs. 95.2% [120/126], respectively), with no significant difference (p > 0.05).
CONCLUSION: 2-VST of ABUS achieved comparable scan coverage and diagnostic performance to that of conventional 3-VST in women with small breasts.
引用
- Kwon BR, Chang JM, Kim SY, et al. Automated Breast Ultrasound System for Breast Cancer Evaluation: Diagnostic Performance of the Two-View Scan Technique in Women with Small Breasts. Korean J Radiol. 2020. PMID: 31920026. doi:10.3348/kjr.2019.0275
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。