同步辐射传播相位对比CT与常规锥形束乳腺CT的乳腺成像比较
研究对8例离体乳腺组织样本分别行同步辐射传播相位对比CT(PB-CT,标准剂量5.8 mGy与低剂量1.5 mGy)和常规锥形束乳腺CT(CBBCT,标准剂量5.8 mGy),由6名医学影像专家按六项图像质量标准评估。标准剂量与低剂量PB-CT在五项质量标准上均显著优于标准剂量CBBCT,提示同步辐射PB-CT可在更低辐射剂量下获得更高图像质量。
要点
- 同步辐射PB-CT在整体质量、可感知对比度、病灶锐利度、正常组织界面及钙化可见性上优于常规CBBCT。
- PB-CT低剂量1.5 mGy即可在多项质量标准上显著优于标准剂量5.8 mGy的CBBCT。
- 观察者间一致性除图像噪声为低可靠性(ICC 0.307)外,其余五项标准为中等可靠性(ICC 0.488–0.633)。
关键影像表现/征象
- 标准剂量PB-CT在整体质量、可感知对比度、病灶锐利度、正常组织界面和钙化可见性五项标准上AUCVGC为0.958至1(p ≤ .05),显著高于标准剂量CBBCT。
- 低剂量PB-CT在同样五项标准上AUCVGC为0.785至0.834(p ≤ .05),显著高于标准剂量CBBCT。
诊断标准与关键数值
- 图像质量比较采用视觉分级特征曲线下面积(AUCVGC),标准剂量PB-CT五项标准AUCVGC 0.958至1(p ≤ .05)。
- 低剂量PB-CT五项标准AUCVGC 0.785至0.834(p ≤ .05)。
- 观察者间一致性用组内相关系数(ICC)评估,五项标准ICC 0.488至0.633为中等可靠性,图像噪声ICC 0.307为低可靠性。
临床应用与阅片要点
- 作者指出同步辐射PB-CT可帮助临床医生诊断乳腺癌患者。
- 阅片时注意图像噪声的观察者间一致性较低(ICC 0.307),对噪声相关评价需谨慎。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate and compare the image quality of propagation-based phase-contrast computed tomography (PB-CT) using synchrotron radiation and conventional cone-beam breast computed tomography (CBBCT) based on various radiological image quality criteria.
METHODS: Eight excised breast tissue samples of various sizes and containing different lesion types were scanned using PB-CT at a synchrotron facility and using CBBCT at a university-affiliated breast imaging centre. PB-CT scans were performed at two different mean glandular dose (MGD) levels: standard (5.8 mGy) and low (1.5 mGy), for comparison with CBBCT scans at the standard MGD (5.8 mGy). Image quality assessment was carried out using six quality criteria and six independent medical imaging experts in a reading room with mammography workstations. The interobserver agreement between readers was evaluated using intraclass correlation coefficient (ICC), and image quality was compared between the two breast imaging modalities using the area under the visual grading characteristic curve (AUCVGC).
RESULTS: Interobserver agreement between the readers showed moderate reliability for five image criteria (ICC: ranging from 0.488 to 0.633) and low reliability for one criterion (image noise) (ICC 0.307). For five image quality criteria (overall quality, perceptible contrast, lesion sharpness, normal tissue interfaces, and calcification visibility), both standard-dose PB-CT images (AUCVGC 0.958 to 1, p ≤ .05) and low dose PB-CT images (AUCVGC 0.785 to 0.834, p ≤ .05) were of significantly higher image quality than standard-dose CBBCT images.
CONCLUSIONS: Synchrotron-based PB-CT can achieve a significantly higher radiological image quality at a substantially lower radiation dose compared with conventional CBBCT.
KEY POINTS: • PB-CT using synchrotron radiation results in higher image quality than conventional CBBCT for breast imaging. • PB-CT using synchrotron radiation requires a lower radiation dose than conventional CBBCT for breast imaging. • PB-CT can help clinicians diagnose patients with breast cancer.