温度敏感型冷冻组织成像用于STIR-UTE MRI监测冷冻消融
本研究开发了短tau反转恢复超短回波时间(STIR-UTE)MRI序列,用于界定冷冻消融中温度低于-40°C的致死性冷冻组织区域。体外实验显示该序列可将-40°C至-8°C区域显示为高信号,临床12例MR引导前列腺癌冷冻消融中STIR-UTE frozen-tissue体积Vinner和Vouter显著小于TSE信号空洞,Vinner与冰球比值0.92±0.08,Vouter为0.29±0.07,单例随访中Vinner与坏死体积强相关。
要点
- STIR-UTE序列利用冷冻组织短T1和T2*特性,将-40°C至-8°C等温区显示为高信号。
- 临床冷冻消融中可见以冷冻探针为中心的暗区Vinner(可能低于-40°C),外周环形高信号区及外层暗区。
- STIR-UTE所测Vinner和Vouter体积显著小于TSE信号空洞即冰球(P < 1.0 × 10)。
关键影像表现/征象
- 离体牛肌肉实验显示STIR-UTE图像将约-40°C至-8°C区域显示为高信号,更低和更高温度组织呈暗信号。
- 临床病例中STIR-UTE图像显示以冷冻探针杆为中心的暗体积Vinner(温度可能低于-40°C),周围环绕 doughnut 形高信号体积(温度可能在-40°C至-8°C之间),外层为暗体积(温度可能高于-8°C),可计算Vouter。
诊断标准与关键数值
- STIR-UTE序列以T1约271毫秒为阈值,因3特斯拉磁场下-40°C至-8°C冷冻组织T1短于该阈值而呈高信号。
- STIR-UTE测量的Vinner和Vouter体积与冰球(TSE信号空洞)比值分别为0.92±0.08和0.29±0.07。
- STIR-UTE frozen-tissue体积Vinner和Vouter显著小于TSE信号空洞(P < 1.0 × 10)。
临床应用与阅片要点
- STIR-UTE可用于MR引导冷冻消融中识别可能高于致死温度的区域,作为术中验证肿瘤消融完成的辅助手段。
局限性
- 单例消融后随访病例中观察到Vinner与坏死体积强相关。
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The aim of this study was to develop a method to delineate the lethally frozen-tissue region (temperature less than -40°C) arising from interventional cryoablation procedures using a short tau inversion-recovery ultrashort echo-time (STIR-UTE) magnetic resonance (MR) imaging sequence. This method could serve as an intraprocedural validation of the completion of tumor ablation, reducing the number of local recurrences after cryoablation procedures.
MATERIALS AND METHODS: The method relies on the short T1 and T2* relaxation times of frozen soft tissue. Pointwise Encoding Time with Radial Acquisition, a 3-dimensional UTE sequence with TE = 70 microseconds, was optimized with STIR to null tissues with a T1 of approximately 271 milliseconds, the threshold T1. Because the T1 relaxation time of frozen tissue in the temperature range of -40°C < temperature < -8°C is shorter than the threshold T1 at the 3-tesla magnetic field, tissues in this range should appear hyperintense. The sequence was evaluated in ex vivo frozen tissue, where image intensity and actual tissue temperatures, measured by thermocouples, were correlated. Thereafter, the sequence was evaluated clinically in 12 MR-guided prostate cancer cryoablations, where MR-compatible cryoprobes were used to destroy cancerous tissue and preserve surrounding normal tissue.
RESULTS: The ex vivo experiment using a bovine muscle demonstrated that STIR-UTE images showed regions approximately between -40°C and -8°C as hyperintense, with tissues at lower and higher temperatures appearing dark, making it possible to identify the region likely to be above the lethal temperature inside the frozen tissue. In the clinical cases, the STIR-UTE images showed a dark volume centered on the cryoprobe shaft, Vinner, where the temperature is likely below -40°C, surrounded by a doughnut-shaped hyperintense volume, where the temperature is likely between -40°C and -8°C. The hyperintense region was itself surrounded by a dark volume, where the temperature is likely above -8°C, permitting calculation of Vouter. The STIR-UTE frozen-tissue volumes, Vinner and Vouter, appeared significantly smaller than signal voids on turbo spin echo images (P < 1.0 × 10), which are currently used to quantify the frozen-tissue volume ("the iceball"). The ratios of the Vinner and Vouter volumes to the iceball were 0.92 ± 0.08 and 0.29 ± 0.07, respectively. In a single postablation follow-up case, a strong correlation was seen between Vinner and the necrotic volume.
CONCLUSIONS: Short tau inversion-recovery ultrashort echo-time MR imaging successfully delineated the area approximately between -40°C and -8°C isotherms in the frozen tissue, demonstrating its potential to monitor the lethal ablation volume during MR-guided cryoablation.
引用
- Tokuda J, Wang Q, Tuncali K, et al. Temperature-Sensitive Frozen-Tissue Imaging for Cryoablation Monitoring Using STIR-UTE MRI. Invest Radiol. 2020. PMID: 31977600. doi:10.1097/RLI.0000000000000642
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。