高强度聚焦超声手术中热损伤预测应考虑组织热参数的温度依赖性

Temperature Dependence of Tissue Thermal Parameters Should Be Considered in the Thermal Lesion Prediction in High-Intensity Focused Ultrasound Surgery.
PubMed 文献 · Ultrasound Med Biol · 2020-01-23

本研究在高强度聚焦超声(HIFU)温度场预测中引入温度依赖性热参数(比热容、热扩散率、热导率),用KZK方程耦合生物传热方程进行数值模拟,并以组织仿体热损伤实验验证。结果显示,对焦域声强3000 W/cm²照射10秒的肝组织,考虑温度依赖性后预测焦点温升较不考虑时低23%、热损伤面积小41%,提示预测HIFU温升时应考虑热参数的温度依赖性,以避免降低HIFU手术的超声输出设置。

要点

  • 预测HIFU肝组织温升时,考虑热参数温度依赖性可使焦点温升预测值较不考虑时低23%。
  • 考虑热参数温度依赖性后,预测热损伤面积较不考虑时小41%。
  • 预测结果经组织仿体中可视化的热损伤实验观察验证。

核心结论

  • 考虑热参数温度依赖性后,HIFU所致焦点温升预测值较不考虑时低23%、热损伤面积小41%。
  • 作者认为预测HIFU诱导的温升时应考虑温度依赖性热参数,以避免降低HIFU手术的超声输出设置。

研究设计与样本

  • 数值模拟研究,采用非线性Khokhlov-Zabolotskaya-Kuznetsov(KZK)方程耦合生物传热(BHT)方程的热声算法。
  • 肝组织热参数依据实验数据设为温度的函数并纳入BHT方程。
  • 以组织仿体中形成的热损伤进行实验验证,并与预测结果比较。

关键影像表现/征象

  • 对焦域声强3000 W/cm²、照射10秒的肝组织,考虑温度依赖性热参数时预测焦点温升较不考虑时低23%。
  • 相同条件下预测热损伤面积较不考虑温度依赖性时小41%。
  • 组织仿体中可视化的热损伤实验观察验证了上述预测。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

This study considers the temperature-dependent thermal parameters (specific heat capacity, thermal diffusivity and thermal conductivity) used when predicting the temperature rise of tissue exposed to high-intensity focused ultrasound (HIFU). Numerical analysis was performed using the Khokhlov-Zabolotskaya-Kuznetsov equation coupled with a bioheat transfer function. The thermal parameters were set as the functions of temperature using experimental data. The results revealed that, for liver tissue exposed to HIFU with a focal intensity of 3000 W/cm2 for 10 s, the predicted focal temperature rise was 23% lower and the thermal lesion area 41% smaller than those predicted without considering the temperature dependence. The prediction was validated by experimental observations on thermal lesions visualized in a tissue-mimicking phantom. The present results suggest that temperature-dependent thermal parameters should be considered in the prediction of HIFU-induced temperature rise to avoid lowering ultrasonic output settings for HIFU surgery. The aim of the present study was to investigate how significantly the temperature dependence of the thermal parameters affects the thermal dose imposed on the tissue by a typical clinical HIFU device. A numerical simulation was performed using a thermo-acoustic algorithm coupling the non-linear Khokhlov-Zabolotskaya-Kuznetsov (KZK) equation (Meaney et al. 1998; Filonenko and Khokhlova 2001) and a bio-heat transfer (BHT) equation (Pennes 1948). Thermal parameters of liver tissue were modeled in the present study as functions of temperature and were incorporated into the BHT equation to compensate for the variations in thermal parameters with temperature. Experimental validation was achieved by comparing the predictions with the thermal lesions formed in the tissue-mimicking phantoms.