冠状动脉CT血管成像对比剂注射方案:对比剂用量应按体重还是体表面积调整
一项前瞻性随机研究将489例疑似冠心病患者分配至按体重调整碘流率方案(245例)或按体表面积调整碘剂量方案(244例)行冠状动脉CT血管成像。两组平均动脉强化、图像噪声、信噪比、对比噪声比及图像质量评分均无统计学显著差异,体重和体表面积与主动脉强化均无相关性。
要点
- 按体重与按体表面积调整对比剂方案在冠状动脉CT血管成像中获得的平均动脉强化和图像质量相当。
- 体重方案组与体表面积方案组的平均动脉强化分别为396.8±47.6 HU和395.8±42.2 HU,差异无统计学意义(p=0.804)。
- 本研究中体重和体表面积均与主动脉强化无相关性(p=0.324和0.932)。
诊断标准与关键数值
- 体重方案组按体重分层设定个体化碘流率:≤50 kg为1 g/s,51–60 kg为1.2 g/s,61–70 kg为1.4 g/s,71–80 kg为1.6 g/s,81–90 kg为1.8 g/s,91–100 kg为2 g/s,>100 kg为2.2 g/s。
- 体表面积方案组在12秒内给予9,600 mg碘/m²对比剂。
- 体重方案组与体表面积方案组的平均动脉强化分别为396.8±47.6 HU和395.8±42.2 HU(p=0.804;95%CI:-7至9)。
- 体重方案组与体表面积方案组的图像噪声分别为25.2±5.8 HU和25.5±5.4 HU(p=0.549),信噪比分别为16.7±4.4和16.6±3.6(p=0.902),对比噪声比分别为25.1±5.8和25.8±7.4(p=0.258),图像质量评分分别为4.1±0.9和4±0.9(p=0.770)。
- 体重与主动脉强化无相关性(p=0.324),体表面积与主动脉强化亦无相关性(p=0.932)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To compare the uniformity and image quality between contrast media injection protocols adjusted for patient body weight (BW) versus body surface area (BSA) during coronary computed tomography (CT) angiography (CCTA).
MATERIALS AND METHODS: Consecutive patients (n=489) with suspected coronary artery disease were randomised prospectively to one of two CCTA protocols. In the BW protocol (n=245), patients received individualised iodine delivery rates (≤50 kg: 1 g/s; 51-60 kg: 1.2 g/s; 61-70 kg: 1.4 g/s; 71-80 kg: 1.6 g/s; 81-90 kg: 1.8 g/s; 91-100 kg: 2 g/s; >100 kg: 2.2 g/s). In the BSA protocol (n=244), patients received 9,600 mg iodine/m2 of contrast medium over 12 seconds. Attenuation and image noise were measured. Signal-to-noise ratio and contrast-to-noise ratio were calculated. Image quality was scored. Attenuation was assessed for correlation with BW and BSA using linear regression.
RESULTS: There were no statistically significant differences in mean arterial attenuation (396.8±47.6 versus 395.8±42.2 HU, p=0.804; 95% confidence interval: -7 to 9), image noise (25.2±5.8 versus 25.5±5.4 HU; p=0.549), signal-to-noise ratio (16.7±4.4 versus 16.6±3.6; p=0.902), contrast-to-noise ratio (25.1±5.8 versus 25.8±7.4; p=0.258) or image quality scores (4.1±0.9 versus 4±0.9; p=0.770) between the BW and BSA protocols. There was no correlation between BW and aortic attenuation or between BSA and aortic attenuation (p=0.324 and 0.932, respectively).
CONCLUSION: The average contrast media attenuation and image quality was comparable between BW-adjusted protocol and BSA-adjusted protocol.
引用
- Yin WH, Yu YT, Zhang Y, et al. Contrast medium injection protocols for coronary CT angiography: should contrast medium volumes be tailored to body weight or body surface area? Clin Radiol. 2020. PMID: 31924327. doi:10.1016/j.crad.2019.12.005