第三代双源CT动态心肌灌注成像的心肌覆盖范围与辐射剂量
回顾性分析107例行动态心肌CT灌注成像患者,比较第二代双源CT 80 kV(54例)与第三代双源CT 70 kV(53例)的图像质量、心肌血流量和辐射剂量。第三代双源CT有效辐射剂量更低(3.97 mSv vs 5.49 mSv,p<0.001),覆盖不全更少(1.9% vs 56%,p<0.001),对比噪声比与心肌血流量无显著差异。
要点
- 第三代双源CT 70 kV动态心肌灌注成像的有效辐射剂量低于第二代双源CT 80 kV(3.97 mSv vs 5.49 mSv)。
- 第三代双源CT 10.5 cm z轴覆盖范围覆盖不全发生率低于第二代双源CT(1.9% vs 56%)。
- 两组对比噪声比及缺血与非缺血心肌血流量均无显著差异。
核心结论
- 第三代双源CT增宽z轴覆盖所致的辐射增加可通过使用70 kV管电压平衡,且不损害图像质量或心肌血流量定量。
- 动态心肌CT灌注成像中10.5 cm的z轴覆盖范围足以在大多数患者中实现左心室心肌的完整覆盖。
研究设计与样本
- 回顾性分析107例行动态心肌CT灌注成像的患者,其中第二代双源CT 80 kV组54例,第三代双源CT 70 kV组53例。
- 比较两组图像质量、估算心肌血流量、辐射剂量及左心室心肌覆盖范围。
- 采用倾向性评分匹配队列比较心肌血流量。
关键影像表现/征象
- 第三代双源CT与第二代双源CT的对比噪声比无显著差异(37.4 ± 11.4 vs 35.5 ± 11.2,p = 0.396)。
- 第二代双源CT覆盖不全发生率高于第三代双源CT(56% [30/54] vs 1.9% [1/53],p < 0.001)。
诊断标准与关键数值
- 第三代双源CT有效辐射剂量为3.97 ± 0.92 mSv(转换因子0.017 mSv/mGy·cm),第二代双源CT为5.49 ± 1.36 mSv(p < 0.001)。
- 倾向性评分匹配队列中,非缺血心肌血流量第三代双源CT为146.2 ± 26.5 mL/min/100 g,第二代双源CT为157.5 ± 34.9 mL/min/100 g(p = 0.137)。
- 倾向性评分匹配队列中,缺血心肌血流量第三代双源CT为92.7 ± 21.1 mL/min/100 g,第二代双源CT为90.9 ± 29.7 mL/min/100 g(p = 0.876)。
临床应用与阅片要点
- 第三代双源CT采用70 kV管电压行动态心肌CT灌注成像可在降低辐射剂量的同时保持图像质量和心肌血流量定量能力。
- 10.5 cm z轴覆盖范围在大多数患者中可实现左心室心肌完整覆盖。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Third-generation dual-source computed tomography (3rd-DSCT) allows dynamic myocardial CT perfusion imaging (dynamic CTP) with a 10.5-cm z-axis coverage. Although the increased radiation exposure associated with the 50% wider scan range compared to second-generation DSCT (2nd-DSCT) may be suppressed by using a tube voltage of 70 kV, it remains unclear whether image quality and the ability to quantify myocardial blood flow (MBF) can be maintained under these conditions. This study aimed to compare the image quality, estimated MBF, and radiation dose of dynamic CTP between 2nd-DSCT and 3rd-DSCT and to evaluate whether a 10.5-cm coverage is suitable for dynamic CTP.
MATERIALS AND METHODS: We retrospectively analyzed 107 patients who underwent dynamic CTP using 2nd-DSCT at 80 kV (n = 54) or 3rd-DSCT at 70 kV (n = 53). Image quality, estimated MBF, radiation dose, and coverage of left ventricular (LV) myocardium were compared.
RESULTS: No significant differences were observed between 3rd-DSCT and 2nd-DSCT in contrast-to-noise ratio (37.4 ± 11.4 vs. 35.5 ± 11.2, p = 0.396). Effective radiation dose was lower with 3rd-DSCT (3.97 ± 0.92 mSv with a conversion factor of 0.017 mSv/mGy·cm) compared to 2nd-DSCT (5.49 ± 1.36 mSv, p < 0.001). Incomplete coverage was more frequent with 2nd-DSCT than with 3rd-DSCT (1.9% [1/53] vs. 56% [30/54], p < 0.001). In propensity score-matched cohorts, MBF was comparable between 3rd-DSCT and 2nd-DSCT in non-ischemic (146.2 ± 26.5 vs. 157.5 ± 34.9 mL/min/100 g, p = 0.137) as well as ischemic myocardium (92.7 ± 21.1 vs. 90.9 ± 29.7 mL/min/100 g, p = 0.876).
CONCLUSION: The radiation increase inherent to the widened z-axis coverage in 3rd-DSCT can be balanced by using a tube voltage of 70 kV without compromising image quality or MBF quantification. In dynamic CTP, a z-axis coverage of 10.5 cm is sufficient to achieve complete coverage of the LV myocardium in most patients.
引用
- Takafuji M, Kitagawa K, Ishida M, et al. Myocardial Coverage and Radiation Dose in Dynamic Myocardial Perfusion Imaging Using Third-Generation Dual-Source CT. Korean J Radiol. 2020. PMID: 31920029. doi:10.3348/kjr.2019.0323
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。