左心室体模中的四维超声粒子图像测速

4-D Echo-Particle Image Velocimetry in a Left Ventricular Phantom.
PubMed 文献 · Ultrasound Med Biol · 2020-01-08

本研究在动态左心室体模中,用原型矩阵经食管探头行高帧率四维超声粒子图像测速(echo-PIV),以光学时间分辨断层PIV(tomo-PIV)为参考标准。两者在左心室血流总体形态上一致,但echo-PIV平滑掉较小血流结构,并在较大成像深度处低估流量。四维echo-PIV可在四个心动周期内完成,但临床转化前仍需解决低空间分辨率区域高速度梯度的分辨方法。

要点

  • echo-PIV与tomo-PIV在左心室血流总体形态上一致,但echo-PIV平滑掉较小血流结构。
  • echo-PIV在较大成像深度处低估流量,与PIV核尺寸及探头点扩散函数相对速度梯度较大有关。
  • 四维echo-PIV可在四个心动周期内完成,仅需短时间屏气。

核心结论

  • 四维echo-PIV与tomo-PIV在左心室血流总体形态上一致,但echo-PIV平滑掉较小血流结构,并在较大成像深度处低估流量。
  • 四维echo-PIV可在四个心动周期内完成,仅需短时间屏气,结果有前景;但临床转化前仍需解决低空间分辨率区域高速度梯度的分辨方法。

研究设计与样本

  • 研究在动态左心室体模中评估高帧率四维超声粒子图像测速(echo-PIV),使用原型矩阵经食管探头。
  • 以光学时间分辨断层PIV(tomo-PIV)作为参考标准进行比较。

关键影像表现/征象

  • echo-PIV与tomo-PIV在左心室血流总体形态上一致,但echo-PIV平滑掉较小血流结构。
  • echo-PIV在较大成像深度处低估流量,该处PIV核尺寸和探头点扩散函数相对速度梯度较大。

临床应用与阅片要点

  • 四维echo-PIV可在四个心动周期内完成,仅需短时间屏气。
  • 临床转化前仍需解决低空间分辨率区域高速度梯度的分辨方法。

局限性

  • echo-PIV平滑掉较小血流结构。
  • echo-PIV在较大成像深度处低估流量。
  • 临床转化前仍需解决低空间分辨率区域高速度梯度的分辨方法。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Left ventricular (LV) blood flow is an inherently complex time-varying 3-D phenomenon, where 2-D quantification often ignores the effect of out-of-plane motion. In this study, we describe high frame rate 4-D echocardiographic particle image velocimetry (echo-PIV) using a prototype matrix transesophageal transducer and a dynamic LV phantom for testing the accuracy of echo-PIV in the presence of complex flow patterns. Optical time-resolved tomographic PIV (tomo-PIV) was used as a reference standard for comparison. Echo-PIV and tomo-PIV agreed on the general profile of the LV flow patterns, but echo-PIV smoothed out the smaller flow structures. Echo-PIV also underestimated the flow rates at greater imaging depths, where the PIV kernel size and transducer point spread function were large relative to the velocity gradients. We demonstrate that 4-D echo-PIV could be performed in just four heart cycles, which would require only a short breath-hold, providing promising results. However, methods for resolving high velocity gradients in regions of poor spatial resolution are required before clinical translation.