1.5 T下非笛卡尔螺旋三维时间飞跃法磁共振血管成像的超快颅内血管成像:体外与健康志愿者临床研究

Ultrafast Intracranial Vessel Imaging With Non-Cartesian Spiral 3-Dimensional Time-of-Flight Magnetic Resonance Angiography at 1.5 T: An In Vitro and Clinical Study in Healthy Volunteers.
PubMed 文献 · Invest Radiol · 2020-01-01

该研究在1.5 T下比较了5种三维时间飞跃法磁共振血管成像序列(包括3种螺旋序列、1种压缩感知加速序列和传统SENSE序列)的颅内血管成像质量。体外实验显示各序列测量血管横截面积无显著差异;体内实验显示螺旋序列在多个定性指标上优于SENSE,扫描时间可缩短至1分32秒(较SENSE减少70.12%),但对比噪声比降低。

要点

  • 螺旋TOF序列可在1分32秒完成颅内血管成像,较SENSE的5分08秒减少70.12%扫描时间。
  • 螺旋序列在多个定性图像质量指标上优于SENSE,但对比噪声比显著低于SENSE。
  • 螺旋序列血管锐度优于SENSE,压缩感知序列与SENSE相当。

关键影像表现/征象

  • 体外实验中,5种TOF序列与质子密度加权序列在测量血管节段横截面积方面无显著差异(P=0.904)。
  • 体内定性评估中,CS 3.5与SENSE定性图像质量相当,而3种螺旋TOF序列在多个类别上优于SENSE(P值从0.002到0.031)。
  • 螺旋0.8和螺旋0.6序列在12个类别中得分显著更高。

诊断标准与关键数值

  • 对比噪声比(CNR,均值±标准差):SENSE 37.48±7.13,CS 3.5 31.14±5.97,螺旋序列19.77±1.65,螺旋0.8 16.18±2.14,螺旋0.6 10.37±1.05;CS 3.5和所有3种螺旋TOF的CNR均显著低于SENSE。
  • 对比率(CR):SENSE 0.53±0.03,螺旋序列0.56±0.03,螺旋序列CR显著改善;其余TOF与SENSE无显著差异。
  • 血管锐度:SENSE 0.37±0.03,螺旋序列0.52±0.07,螺旋0.8 0.53±0.08,螺旋0.6 0.73±0.09;所有螺旋TOF均优于SENSE,CS 3.5与SENSE表现相当(0.37±0.03)。
  • 半高全宽值在任何TOF序列间均无显著差异。

临床应用与阅片要点

  • 螺旋TOF序列可能以超快扫描时间(短至1分32秒)提供与SENSE相当或部分定性指标更优的颅内血管成像质量。
  • 螺旋TOF图像的对比噪声比降低,阅片时需注意此差异。
  • 螺旋成像可选用于提高空间分辨率同时保持笛卡尔采集的扫描时间。

局限性

  • 读者间一致性范围从差(alpha=-0.013,颈内动脉C7段显示)到中等(alpha=0.737,矢状位可见血管数量)。
  • 螺旋TOF图像的对比噪声比降低。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Non-Cartesian spiral magnetic resonance (MR) acquisition may enable higher scan speeds, as the spiral traverses the k-space more efficiently per given time than in Cartesian trajectories. Spiral MR imaging can be implemented in time-of-flight (TOF) MR angiography (MRA) sequences. In this study, we tested the performance of five 3-dimensional TOF MRA sequences for intracranial vessel imaging at 1.5 T with qualitative and quantitative image quality metrics based on in vitro and in vivo measurements. Specifically, 3 novel spiral TOF MRA sequences (spiral-TOFs) and a compressed sensing (CS) technology-accelerated TOF MRA sequence (CS 3.5) were compared with a conventional (criterion standard) parallel imaging-accelerated TOF MRA sequence (SENSE).

MATERIALS AND METHODS: The SENSE sequence (5:08 minutes) was compared with the CS 3.5 sequence (3:06 minutes) and a spiral-TOF (spiral, 1:32 minutes), all with identical resolutions. In addition, 2 further isotropic spiral-TOFs (spiral 0.8, 2:12 minutes; spiral 0.6, 5:22 minutes) with higher resolution were compared with the SENSE. First, vessel tracking experiments were performed in vitro with a dedicated vascular phantom to determine possible differences in the depiction of cross-sectional areas of vessel segments. For the in vitro tests, an additional 3-dimensional proton density-weighted sequence was added for comparison reasons. Second, 3 readers blinded to sequence details assessed qualitative (16 features) and 2 readers assessed quantitative (contrast-to-noise ratio [CNR], contrast ratio [CR], vessel sharpness, and full width at half maximum edge criterion measurements) image quality based on images acquired from scanning 10 healthy volunteers with all 5 TOF sequences. Scores from quantitative image quality analysis were compared with Kruskal-Wallis, analysis of variance, or Welch's analysis of variance, followed by Dunnett's or Dunnett's T3 post hoc tests. Scores from qualitative image quality analysis were compared with exact binomial tests, and the level of interreader agreement was determined with Krippendorff's alpha.

RESULTS: Concerning the in vitro tests, there were no significant differences between the 5 TOFs and the proton density-weighted sequence in measuring cross-sectional areas of vessel segments (P = 0.904). As for the in vivo tests, the CS 3.5 exhibited equal qualitative image quality as the SENSE, whereas the 3 spiral-TOFs outperformed the SENSE in several categories (P values from 0.002 to 0.031). Specifically, the spiral 0.8 and 0.6 sequences achieved significantly higher scores in 12 categories. Interreader agreement ranged from poor (alpha = -0.013, visualization of internal carotid artery segment C7) to substantial (alpha = 0.737, number of vessels visible, sagittal). As for the quantitative metrics, the CS 3.5 and all 3 spiral-TOFs presented with significantly worse CNR than the SENSE ([mean ± SD] SENSE 37.48 ± 7.13 vs CS 3.5 31.14 ± 5.97 vs spiral 19.77 ± 1.65 vs spiral 0.8 16.18 ± 2.14 vs spiral 0.6 10.37 ± 1.05). The CR values did not differ significantly between the SENSE and the other TOFs except for the spiral sequence that showed significantly improved CR (SENSE 0.53 ± 0.03 vs spiral 0.56 ± 0.03). As for vessel sharpness, the SENSE was outperformed by all spiral-TOFs (SENSE 0.37 ± 0.03 vs spiral 0.52 ± 0.07 vs spiral 0.8 0.53 ± 0.08 vs spiral 0.6 0.73 ± 0.09), whereas the CS 3.5 performed equally well (SENSE 0.37 ± 0.03 vs CS 3.5 0.37 ± 0.03). Full width at half maximum values did not differ significantly between any TOF.

CONCLUSIONS: Spiral-TOFs may deliver high-quality intracranial vessel imaging thus matching the performance of conventional parallel imaging-accelerated TOFs (such as the SENSE). Specifically, imaging can be performed at unprecedented scan times as short as 1:32 minutes per sequence (70.12% scan time reduction compared with SENSE). Optionally, spiral imaging may also be used to increase spatial resolution while maintaining the scan time of a Cartesian-based acquisition schema. The CNR was decreased in spiral-TOF images.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Sartoretti T, van Smoorenburg L, Sartoretti E, et al. Ultrafast Intracranial Vessel Imaging With Non-Cartesian Spiral 3-Dimensional Time-of-Flight Magnetic Resonance Angiography at 1.5 T: An In Vitro and Clinical Study in Healthy Volunteers. Invest Radiol. 2020. PMID: 31895223. doi:10.1097/RLI.0000000000000641