人体先进单体素1H磁共振波谱技术:专家共识建议
该专家共识针对传统质子MRS在细微代谢物变化及高场(≥3 T)应用中的局限,推荐采用先进MRS方法(含脉冲序列与调节流程),并给出体素定位、B0匀场、射频脉冲校准、水信号采集、数据保存与质控等具体建议,同时向厂商提出开发先进MRS工具的建议。
要点
- 高场(≥3 T)下化学位移位移误差、B0与B1不均匀性及发射B1场限制显著,推荐使用先进MRS方法。
- TE 25-30 ms可接受时推荐半绝热LASER序列;TE需更短时建议半绝热SPECIAL序列作为替代。
- 需优化B0匀场、逐体素校准射频脉冲,采集未抑制水信号用于涡流校正并最好用于代谢物定量。
诊断标准与关键数值
- TE 25-30 ms可接受时推荐半绝热LASER序列;TE需更短时建议半绝热SPECIAL序列作为替代。
- 应优化B0匀场并逐体素校准射频脉冲。
- 应采集未抑制水信号用于涡流校正,并最好也用于代谢物定量。
- 代谢物与水数据应按单次采集保存,以利于相位和频率对齐并排除运动损坏的采集。
- 最终平均谱应评估SNR、线宽、水抑制效率及非期望相干的存在;不满足预设质量标准的谱应排除进一步分析。
临床应用与阅片要点
- 推荐使用商用工具进行体素定位,以在一致解剖位置采集所有数据,尤其适用于纵向研究。
- 为便于MRS社区使用先进方法,提供了主要临床平台上这些先进协议的资源列表。
- 向厂商提出建议以在标准平台上开发先进MRS,包括实现先进定位序列、扫描仪质控工具及前瞻性体积跟踪与动态线性匀场校正工具。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Conventional proton MRS has been successfully utilized to noninvasively assess tissue biochemistry in conditions that result in large changes in metabolite levels. For more challenging applications, namely, in conditions which result in subtle metabolite changes, the limitations of vendor-provided MRS protocols are increasingly recognized, especially when used at high fields (≥3 T) where chemical shift displacement errors, B0 and B1 inhomogeneities and limitations in the transmit B1 field become prominent. To overcome the limitations of conventional MRS protocols at 3 and 7 T, the use of advanced MRS methodology, including pulse sequences and adjustment procedures, is recommended. Specifically, the semiadiabatic LASER sequence is recommended when TE values of 25-30 ms are acceptable, and the semiadiabatic SPECIAL sequence is suggested as an alternative when shorter TE values are critical. The magnetic field B0 homogeneity should be optimized and RF pulses should be calibrated for each voxel. Unsuppressed water signal should be acquired for eddy current correction and preferably also for metabolite quantification. Metabolite and water data should be saved in single shots to facilitate phase and frequency alignment and to exclude motion-corrupted shots. Final averaged spectra should be evaluated for SNR, linewidth, water suppression efficiency and the presence of unwanted coherences. Spectra that do not fit predefined quality criteria should be excluded from further analysis. Commercially available tools to acquire all data in consistent anatomical locations are recommended for voxel prescriptions, in particular in longitudinal studies. To enable the larger MRS community to take advantage of these advanced methods, a list of resources for these advanced protocols on the major clinical platforms is provided. Finally, a set of recommendations are provided for vendors to enable development of advanced MRS on standard platforms, including implementation of advanced localization sequences, tools for quality assurance on the scanner, and tools for prospective volume tracking and dynamic linear shim corrections.
引用
- Öz G, Deelchand DK, Wijnen JP, et al. Advanced single voxel 1 H magnetic resonance spectroscopy techniques in humans: Experts' consensus recommendations. NMR Biomed. 2020. PMID: 31922301. doi:10.1002/nbm.4236
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。