文献速递技术与安全CT

CT深度学习重建:图像特征的体模研究

Deep Learning Reconstruction at CT: Phantom Study of the Image Characteristics.
PubMed 文献 · Acad Radiol · 2020-01-01

本体模研究在320排CT上比较深度学习重建(DLR)与滤波反投影、混合迭代重建、基于模型的迭代重建的图像噪声、空间分辨率和任务可检测性。DLR图像噪声更低,尤其低剂量时;高对比空间分辨率和任务可检测性优于其他先进技术。

要点

  • DLR在低辐射剂量下噪声低于其他重建方法。
  • DLR高对比空间分辨率与任务可检测性优于其他先进技术。
  • 基于模型的迭代重建空间分辨率高于DLR,但低对比物体MTF与DLR相当。

关键影像表现/征象

  • DLR图像噪声低于其他重建图像,尤其在低辐射剂量设置下。
  • 噪声功率谱显示DLR噪声幅度更低,尤其低频成分。
  • 基于MTF,基于模型的迭代重建空间分辨率高于DLR图像。
  • 低对比物体上,DLR的MTF与其他方法重建图像相当。
  • 模型观察者研究显示,在降低辐射剂量设置下DLR可检测性最佳。

诊断标准与关键数值

  • 噪声特征用CT值标准差和噪声功率谱计算。
  • 空间分辨率用10%调制传递函数(MTF)水平评估。
  • 任务可检测性用模型观察者方法评估。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Noise, commonly encountered on computed tomography (CT) images, can impact diagnostic accuracy. To reduce the image noise, we developed a deep-learning reconstruction (DLR) method that integrates deep convolutional neural networks into image reconstruction. In this phantom study, we compared the image noise characteristics, spatial resolution, and task-based detectability on DLR images and images reconstructed with other state-of-the art techniques.

METHODS: We scanned a phantom harboring cylindrical modules with different contrast on a 320-row detector CT scanner. Phantom images were reconstructed with filtered back projection, hybrid iterative reconstruction, model-based iterative reconstruction, and DLR. The standard deviation of the CT number and the noise power spectrum were calculated for noise characterization. The 10% modulation-transfer function (MTF) level was used to evaluate spatial resolution; task-based detectability was assessed using the model observer method.

RESULTS: On images reconstructed with DLR, the noise was lower than on images subjected to other reconstructions, especially at low radiation dose settings. Noise power spectrum measurements also showed that the noise amplitude was lower, especially for low-frequency components, on DLR images. Based on the MTF, spatial resolution was higher on model-based iterative reconstruction image than DLR image, however, for lower-contrast objects, the MTF on DLR images was comparable to images reconstructed with other methods. The machine observer study showed that at reduced radiation-dose settings, DLR yielded the best detectability.

CONCLUSION: On DLR images, the image noise was lower, and high-contrast spatial resolution and task-based detectability were better than on images reconstructed with other state-of-the art techniques. DLR also outperformed other methods with respect to task-based detectability.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Higaki T, Nakamura Y, Zhou J, et al. Deep Learning Reconstruction at CT: Phantom Study of the Image Characteristics. Acad Radiol. 2020. PMID: 31818389. doi:10.1016/j.acra.2019.09.008

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。