锥形束CT引导内镜逆行胰胆管造影中的辐射剂量
研究回顾性收集了42例使用锥形束CT(CBCT)的内镜逆行胰胆管造影(ERCP)操作中的X线辐射剂量数据,并与常规ERCP的辐射暴露数据进行比较。结果显示,常规ERCP的中位总剂量面积乘积为6.52 Gycm²,使用标准DR协议时为48.9 Gycm²,使用低剂量DR care协议时为19.7 Gycm²,常规ERCP的辐射剂量显著低于两种CBCT协议(p < 0.001),但常规ERCP中存在大量剂量偏高甚至超过CBCT操作的离群值。作者认为,对于操作简单、X线剂量小且透视时间短的病例,常规二维X线技术已足够且辐射剂量低于CBCT;CBCT可能有助于处理胆管解剖困难的ERCP病例,可能减少操作时间、X线曝光时间和对比剂注射需求。
要点
- 常规ERCP中位总剂量面积乘积为6.52 Gycm²,显著低于标准DR协议(48.9 Gycm²)和低剂量DR care协议(19.7 Gycm²)(p < 0.001)。
- 单次标准DR容积扫描中位剂量面积乘积为24.4 Gycm²,单次DR care容积扫描为5.07 Gycm²。
- 常规ERCP虽总体剂量较低,但存在大量剂量偏高甚至超过CBCT操作的离群值。
诊断标准与关键数值
- 单次标准DR容积扫描中位剂量面积乘积为24.4 Gycm²,单次DR care容积扫描中位剂量面积乘积为5.07 Gycm²。
- 常规ERCP中位总剂量面积乘积为6.52 Gycm²,使用DR协议的操作中位总剂量面积乘积为48.9 Gycm²,使用DR care协议的操作中位总剂量面积乘积为19.7 Gycm²。
- 常规ERCP的辐射剂量显著低于使用任一CBCT协议的操作(p < 0.001)。
局限性
- 原文为回顾性数据收集。
- 常规ERCP虽总体剂量较低,但存在大量剂量偏高甚至超过CBCT操作的离群值,剂量范围宽。
- 需要预测高辐射ERCP的方法,以标准化和优化CBCT-ERCP的患者选择。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate the practical and radiation safety considerations in utilization of 3-D imaging through cone beam computed tomography (CBCT) in endoscopic retrograde cholangiopancreatography (ERCP).
METHOD: ERCP procedures were performed using an Artis Q interventional suite (Siemens Healthineers), including 42 procedures with CBCT. Some CBCT cases used the standard "DR" exposure protocol and some used the low dose protocol "DR care". Data on x-ray radiation doses were retrospectively collected and to help optimize technical factors surrounding the procedure and compared to radiation exposure data of patients undergoing conventional ERCP.
RESULTS: The median dose area product was 24.4 Gycm2 for one DR volume and 5.07 Gycm2 for one DR care volume. The median total dose area product was 6.52 Gycm2 for conventional ERCP, 48.9 Gycm2 for procedures using DR and 19.7 Gycm2 when using DR care. Conventional ERCP resulted in a significantly lower radiation dose than procedures using either CBCT protocols (p < 0.001). However, conventional ERCP showed a large number of outliers with higher dose at the level of, or surpassing, CBCT procedures.
CONCLUSIONS: For less complex procedures with small x-ray doses and short fluoroscopy times conventional 2-D x-ray technique is clinically sufficient and utilizes less radiation dose than CBCT methods, but there is a wide range of doses and extreme outliers have been observed. CBCT can facilitate ERCP in cases with difficult ducal anatomy, possibly reducing procedural time, x-ray exposure time and need for contrast injections. Methods for prediction on high radiation ERCPs are needed to standardize and optimize patient selection for CBCT-ERCP.