单源双能CT检测骨髓病变:迭代重建与算法的影响

Single-source dual-energy computed tomography for the detection of bone marrow lesions: impact of iterative reconstruction and algorithms.
PubMed 文献 · Skeletal Radiol · 2019-12-10

本研究以MRI为参考标准,比较单源双能CT不同重建算法对椎体压缩骨折患者骨髓病变(BML)的检测效能。纳入17例患者39个椎体,IR1-IR3在3MD下敏感度87.5%优于FBP的75%,与RD的83.3%相当;所有2MD图像敏感度低于38%。结论认为IR可改善骨髓图像质量,2MD不适用于BML检测。

要点

  • 以MRI为参考标准,3MD结合迭代重建(IR1-IR3)检测BML的敏感度为87.5%,优于FBP的75%。
  • 所有2MD算法的敏感度均低于38%,不适用于BML检测。
  • 3MD-IR1-IR3的图像质量评分显著高于3MD-FBP(p<0.0001)及所有2MD数据集(p<0.03)。

关键影像表现/征象

  • 3MD-IR1-IR3的图像质量评分显著高于3MD-FBP(p<0.0001)及所有2MD数据集(p<0.03)。
  • SNR和CNR测量结果支持上述图像质量评分的模式。

诊断标准与关键数值

  • 以MRI为参考标准,3MD-IR1-IR3检测BML的敏感度为87.5%,3MD-FBP为75%,RD为83.3%,所有2MD图像敏感度低于38%。

临床应用与阅片要点

  • 作者认为,采用迭代重建较FBP可改善单源双能CT骨髓图像质量。
  • 作者认为,基于原始数据的重建较基于图像数据的重建无显著改善。
  • 作者认为,2MD算法不适用于BML检测。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare the diagnostic performance of different reconstruction algorithms of single-source dual-energy computed tomography (DECT) for the detection of bone marrow lesions (BML) in patients with vertebral compression fracture using MRI as the standard of reference.

MATERIAL AND METHODS: Seventeen patients with an age over 50 who underwent single-source DECT of the spine were included. The raw data (RD) were reconstructed using filtered back-projection (FBP) and iterative reconstruction (IR) with three iteration levels (IR1-IR3). Bone marrow images were generated using a three-material decomposition (3MD) and a two-material decomposition (2MD) algorithm and an RD-based approach. Three blinded readers scored the images for image quality and the presence of bone marrow lesions (BML). Only vertebrae with height loss were included. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. The different reconstructions were compared using Dunn's multiple comparison test.

RESULTS: Thirty-nine vertebrae were included. IR(1-3) showed superior sensitivity (87.5%) compared to FBP (75%) using 3MD but was comparable to RD (83.3%). All 2MD images were inferior (sensitivity < 38%). The image quality score was significantly higher for 3MD-IR(1-3) compared to 3MD-FBP (p < 0.0001) and all 2MD data sets (p < 0.03). This pattern was also supported by the SNR and CNR measurements. RD showed no significant improvement compared to IR.

CONCLUSION: The image quality of bone marrow images acquired with DECT can be improved by using IR compared with FBP. RD-based reconstruction does not offer significant improvement over image data-based reconstruction. 2MD algorithms are not suitable for BML detection.