双能光谱CT与定量CT测量骨密度的准确性:体模对比研究

The accuracy of bone mineral density measurement using dual-energy spectral CT and quantitative CT: a comparative phantom study.
PubMed 文献 · Clin Radiol · 2019-12-25

该体模研究以欧洲脊柱体模(ESP)中50、100、200 mg/cm³的羟基磷灰石插入物为对象,分别用光谱CT和QCT协议各扫描10次,比较骨密度测量值与真实值的偏差。两种方法重复性均好(光谱CT变异系数<2.31%,QCT<1.17%),光谱CT在50和100 mg/cm³处的相对误差小于QCT,结论认为两者均可准确测量骨密度,光谱CT偏差更小。

要点

  • 光谱CT与QCT测量骨密度重复性均好,变异系数分别<2.31%与<1.17%。
  • 50与100 mg/cm³插入物处光谱CT相对误差(2.34%、2%)低于QCT(11%、9.85%)。
  • 200 mg/cm³处QCT相对误差4.04%低于光谱CT的5.96%。

核心结论

  • QCT与光谱CT均可准确测量骨密度,光谱CT的偏差小于QCT。

研究设计与样本

  • 体模对比研究,使用含50、100、200 mg/cm³羟基磷灰石插入物的欧洲脊柱体模(ESP),光谱CT与QCT协议各扫描10次。

关键影像表现/征象

  • 采用基于羟基磷灰石的光谱CT物质分解图像和QCT图像测量骨密度值,并与ESP真实值比较。

诊断标准与关键数值

  • 光谱CT相对误差:50 mg/cm³为2.34%(48.83±1.13 mg/cm³),100 mg/cm³为2%(98±1.25 mg/cm³),200 mg/cm³为5.96%(188.09±2.11 mg/cm³)。
  • QCT相对误差:50 mg/cm³为11%(55.5±0.65 mg/cm³),100 mg/cm³为9.85%(109.85±0.42 mg/cm³),200 mg/cm³为4.04%(208.07±0.54 mg/cm³)。
  • 重复性变异系数:光谱CT<2.31%,QCT<1.17%。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To compare bone mineral density (BMD) measurement accuracy by dual-energy spectral computed tomography (CT) and quantitative CT (QCT) using an anthropomorphic European spine phantom (ESP).

MATERIALS AND METHODS: An ESP containing three hydroxyapatite inserts with densities of 50, 100, and 200 mg/cm3 was scanned 10 times each using spectral CT and QCT protocols. Their BMD values were measured using hydroxyapatite-based spectral CT material decomposition images and QCT images and compared with the true values in ESP.

RESULTS: Both protocols had good repeatability in BMD measurement with low coefficient-of-variance (spectral CT: <2.31%; QCT: <1.17%). There were biases in BMD measurement for the 50, 100, and 200 mg/cm3 hydroxyapatite insert with relative errors of 2.34% (48.83±1.13 mg/cm3), 2% (98±1.25 mg/cm3) and 5.96% (188.09±2.11 mg/cm3), in spectral CT, and 11% (55.5±0.65 mg/cm3), 9.85% (109.85±0.42 mg/cm3) and 4.04% (208.07±0.54 mg/cm3) in QCT, respectively.

CONCLUSION: BMD can be accurately measured using either QCT or spectral CT, and spectral CT has smaller bias than QCT.

来源:Clin Radiol,查看原文。版权归原作者所有。