不同类型乳腺X线摄影设备对筛查结果的影响:韩国乳腺癌筛查联盟报告

Effect of Different Types of Mammography Equipment on Screening Outcomes: A Report by the Alliance for Breast Cancer Screening in Korea.
PubMed 文献 · Korean J Radiol · 2019-01-01

该回顾性研究纳入2005至2010年韩国10家医院128756组乳腺X线摄影,比较胶片屏乳腺X线摄影(FSM)、计算机X线摄影乳腺摄影(CRM)与数字乳腺X线摄影(DM)的筛查效能。DM在召回率、癌症检出率、阳性预测值1和特异度上优于FSM和CRM,但敏感度、区间癌率和AUC无显著差异;DM的原位癌检出比例更高。作者认为DM可能有助于提高筛查质量,但需评估原位癌的过度诊断问题。

要点

  • DM的召回率14.8%、癌症检出率3.4/1000、阳性预测值1为2.3%、特异度85.5%,均优于FSM和CRM(p<0.001)。
  • DM、FSM、CRM的AUC分别为0.898、0.860、0.866,差异无统计学意义(p=0.150)。
  • DM的原位癌比例为27.5%,高于CRM的13.6%和FSM的11.8%(p=0.003),需评估过度诊断问题。

核心结论

  • 与FSM和CRM相比,DM在召回率、癌症检出率、阳性预测值1和特异度上表现更好,AUC相似,且DM检出更多原位癌;DM的应用可能有助于提高乳腺X线筛查质量,但需评估DM使用中原位癌的过度诊断问题。

研究设计与样本

  • 回顾性纳入2005至2010年韩国乳腺癌筛查联盟10家医院的128756组乳腺X线摄影,比较FSM、CRM和DM的诊断准确性与筛查效能指标。

诊断标准与关键数值

  • DM的AUC为0.898(95%CI 0.878-0.919),FSM为0.860(95%CI 0.815-0.905),CRM为0.866(95%CI 0.828-0.903),三者比较p=0.150。DM召回率14.8%,FSM 24.8%,CRM 19.8%;DM癌症检出率3.4/1000次检查,FSM 2.2/1000,CRM 2.1/1000;DM阳性预测值1为2.3%,FSM 0.9%,CRM 1.1%;DM特异度85.5%,FSM 75.3%,CRM 80.3%(以上指标p<0.001)。DM敏感度86.3%,FSM 87.4%,CRM 86.3%;DM区间癌率0.6%,FSM 0.4%,CRM 0.4%。DM原位癌比例27.5%,CRM 13.6%,FSM 11.8%(p=0.003)。

临床应用与阅片要点

  • 作者认为DM的应用可能有助于提高乳腺X线筛查质量,但需评估DM使用中原位癌的过度诊断问题。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To investigate the effects of different types of mammography equipment on screening outcomes by comparing the performance of film-screen mammography (FSM), computed radiography mammography (CRM), and digital mammography (DM).

MATERIALS AND METHODS: We retrospectively enrolled 128756 sets of mammograms from 10 hospitals participating in the Alliance for Breast Cancer Screening in Korea between 2005 and 2010. We compared the diagnostic accuracy of the types of mammography equipment by analyzing the area under the receiver operating characteristic curve (AUC) with a 95% confidence interval (CI); performance indicators, including recall rate, cancer detection rate (CDR), positive predictive value₁ (PPV₁), sensitivity, specificity, and interval cancer rate (ICR); and the types of breast cancer pathology.

RESULTS: The AUCs were 0.898 (95% CI, 0.878-0.919) in DM, 0.860 (0.815-0.905) in FSM, and 0.866 (0.828-0.903) in CRM (p = 0.150). DM showed better performance than FSM and CRM in terms of the recall rate (14.8 vs. 24.8 and 19.8%), CDR (3.4 vs. 2.2 and 2.1 per 1000 examinations), PPV₁ (2.3 vs. 0.9 and 1.1%), and specificity (85.5 vs. 75.3 and 80.3%) (p < 0.001) but not in terms of sensitivity (86.3 vs. 87.4 and 86.3%) and ICR (0.6 vs. 0.4 and 0.4). The proportions of carcinoma in situ (CIS) were 27.5%, 13.6%, and 11.8% for DM, CRM, and FSM, respectively (p = 0.003).

CONCLUSION: In comparison to FSM and CRM, DM showed better performance in terms of the recall rate, CDR, PPV₁, and specificity, although the AUCs were similar, and more CISs were detected using DM. The application of DM may help to improve the quality of mammography screenings. However, the overdiagnosis issue of CIS using DM should be evaluated.

来源:Korean J Radiol,查看原文。版权归原作者所有。