多中心DISCHARGE试验预试验:冠状动脉CT血管成像与有创冠状动脉造影的图像质量及方案依从性结果

Pilot study of the multicentre DISCHARGE Trial: image quality and protocol adherence results of computed tomography and invasive coronary angiography.
PubMed 文献 · Eur Radiol · 2019-12-16

在多中心DISCHARGE试验预试验中,回顾性评估各中心CCTA与ICA数据集对方案及ICA推荐标准的依从性、图像质量与非诊断率。226例中2.7%不符合纳入/排除标准,76例ICA中17.1%未达预设标准;经反馈后CCTA非诊断率由21.4%降至12%(p=0.13),信噪比与对比噪声比无显著改善。

要点

  • 预试验中CCTA方案依从性不达标比例高达77.6%。
  • 图像质量在不同中心间存在差异,核心实验室反馈可改善图像质量。
  • CCTA非诊断率由初始数据库21.4%降至最终数据库12%,但未达统计学显著(p=0.13)。

诊断标准与关键数值

  • 226例数据集中6例(2.7%)不符合纳入/排除标准。
  • 76例ICA数据集中13例(17.1%)未达预设标准。
  • 预设标准不达标比例在初始CCTA数据库与最终接受数据库之间下降(多项回归:70例中53例[76%] vs 75例中17例[23%])。
  • 最终接受数据库的信噪比(p=0.20)与对比噪声比(p=0.09)较初始数据库无显著改善。
  • CTA非诊断率:初始CCTA数据库70例中15例(21.4%) vs 最终接受数据库75例中9例(12%),p=0.13,未达统计学显著。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To implement detailed EU cardiac computed tomography angiography (CCTA) quality criteria in the multicentre DISCHARGE trial (FP72007-2013, EC-GA 603266), we reviewed image quality and adherence to CCTA protocol and to the recommendations of invasive coronary angiography (ICA) in a pilot study.

MATERIALS AND METHODS: From every clinical centre, imaging datasets of three patients per arm were assessed for adherence to the inclusion/exclusion criteria of the pilot study, predefined standards for the CCTA protocol and ICA recommendations, image quality and non-diagnostic (NDX) rate. These parameters were compared via multinomial regression and ANOVA. If a site did not reach the minimum quality level, additional datasets had to be sent before entering into the final accepted database (FADB).

RESULTS: We analysed 226 cases (150 CCTA/76 ICA). The inclusion/exclusion criteria were not met by 6 of the 226 (2.7%) datasets. The predefined standard was not met by 13 of 76 ICA datasets (17.1%). This percentage decreased between the initial CCTA database and the FADB (multinomial regression, 53 of 70 vs 17 of 75 [76%] vs [23%]). The signal-to-noise ratio and contrast-to-noise ratio of the FADB did not improve significantly (ANOVA, p = 0.20; p = 0.09). The CTA NDX rate was reduced, but not significantly (initial CCTA database 15 of 70 [21.4%]) and FADB 9 of 75 [12%]; p = 0.13).

CONCLUSION: We were able to increase conformity to the inclusion/exclusion criteria and CCTA protocol, improve image quality and decrease the CCTA NDX rate by implementing EU CCTA quality criteria and ICA recommendations.

KEY POINTS: • Failure to meet protocol adherence in cardiac CTA was high in the pilot study (77.6%). • Image quality varies between sites and can be improved by feedback given by the core lab. • Conformance with new EU cardiac CT quality criteria might render cardiac CTA findings more consistent and comparable.