儿童胸部CT采用100 kVp联合锡滤过与70 kVp在相似辐射剂量下的图像质量比较

Pediatric chest computed tomography at 100 kVp with tin filtration: comparison of image quality with 70-kVp imaging at comparable radiation dose.
PubMed 文献 · Pediatr Radiol · 2019-12-18

研究纳入50例连续儿童(组1)行锡滤过100 kVp非增强胸部CT,并按体重和年龄配对50例70 kVp扫描儿童(组2),两组辐射剂量相当;组1空气和主动脉内客观噪声显著低于组2,平均噪声降低21.6%(胸廓周围)和12.0%(胸腔内),两组肺和纵隔图像主观质量无显著差异。

要点

  • 锡滤过100 kVp胸部CT在相似辐射剂量下较70 kVp方案降低客观噪声。
  • 胸廓周围平均噪声降低21.6%,胸腔内平均噪声降低12.0%。
  • 两组肺和纵隔图像主观质量无显著差异,主观评分均优秀。

关键影像表现/征象

  • 组1空气测量图像噪声显著低于组2(P<0.0001)。
  • 组1主动脉内测量图像噪声显著低于组2(P<0.001)。
  • 组1胸廓周围平均噪声降低21.6%(标准差16.1),胸腔内平均噪声降低12.0%(标准差32.7)。
  • 两组肺和纵隔图像主观质量无显著差异,总体主观评分均优秀。

临床应用与阅片要点

  • 作者认为锡滤过100 kVp在相似辐射剂量下可降低图像噪声,为后续研究在不改变图像质量的前提下降低剂量提供了基础。

局限性

  • 原文结论指出需进一步研究以在不改变图像质量的前提下降低剂量。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Radiation dose reduction is a primary objective in pediatric populations owing to the well-known risks of radiation-induced cancers. Low-energy photons participate in the radiation dose without significantly contributing to image formation. Their suppression by means of tin filtration should decrease the image noise, anticipating a subsequent application to dose saving.

OBJECTIVE: To investigate the level of noise reduction achievable with tin (Sn) filtration at 100 kVp for chest computed tomography (CT) in comparison with a standard scanning mode at 70 kVp with comparable radiation dose.

MATERIALS AND METHODS: Fifty consecutive children (Group 1) underwent non-contrast chest CT examinations on a third-generation dual-source CT system at tin-filtered 100 kVp and pitch 2. The tube-current time product (mAs) was adjusted to maintain the predicted dose length product (DLP) value at 70 kVp for the respective patient. Each child was then paired by weight and age to a child scanned at 70 kVp on the same CT unit (Group 2); Group 2 patients were consecutive patients, retrospectively selected from our database of children prospectively scanned at 70 kVp. Objective and subjective image quality were compared between the two groups of patients to investigate the overall image quality and level of noise reduction that could be subsequently achievable with tin filtration in clinical practice.

RESULTS: The mean image noise was significantly lower in Group 1 compared to Group 2 when measured in the air (P<0.0001) and inside the aorta (P<0.001). The mean noise reduction was 21.6% (standard deviation [SD] 16.1) around the thorax and 12.0% (SD 32.7) inside the thorax. There was no significant difference in subjective image quality of lung and mediastinal images with excellent overall subjective scores in both groups.

CONCLUSION: At comparable radiation dose, the image noise was found to be reduced by 21.6% compared to the 70-kVp protocol, providing basis for dose reduction without altering image quality in further investigations.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Vivier S, Deken V, Arous Y, et al. Pediatric chest computed tomography at 100 kVp with tin filtration: comparison of image quality with 70-kVp imaging at comparable radiation dose. Pediatr Radiol. 2019. PMID: 31853569. doi:10.1007/s00247-019-04543-w