对比剂剂量减半的静脉期胸部CT:利用能谱低keV单能图像改善图像质量
研究纳入30例临床指征下对比剂剂量减少50%的患者,行静脉期双层探测器CT扫描,重建常规多能图像与40 keV单能图像。单能图像在肺门淋巴结对比度、信噪比、对比噪声比及主观图像质量评分上均显著优于多能图像,且无单能检查被判为不可诊断。
要点
- 对比剂减量50%的静脉期胸部双层探测器CT中,40 keV单能图像肺门淋巴结 delineation 与对比增强评分均显著优于常规多能图像。
- 单能图像肺门淋巴结对比度、SNR、CNR均显著高于多能图像,所有单能检查均满足整体对比增强要求。
- 低keV单能重建可补偿对比剂减量带来的图像质量下降,作者认为减量方案可考虑用于更广泛的双层探测器CT检查与患者。
关键影像表现/征象
- 单能图像肺门淋巴结与邻近右肺动脉的对比度差值显著高于多能图像(92.3 ± 37.9 HU vs. 33.1 ± 14.5 HU,p < 0.01)。
- 单能图像信噪比显著高于多能图像(8.4 ± 3.4 vs. 4.0 ± 1.2,p < 0.01)。
- 单能图像对比噪声比显著高于多能图像(9.2 ± 6.3 vs. 2.6 ± 1.5,p < 0.01)。
诊断标准与关键数值
- 淋巴结 delineation 主观评分单能图像显著优于多能图像(3.7 ± 0.9 vs. 1.8 ± 0.7,p < 0.01),采用5点Likert量表。
- 整体对比增强主观评分单能图像显著优于多能图像(3.9 ± 0.7 vs. 2.3 ± 0.7,p < 0.01),采用5点Likert量表。
- 无单能检查被分类为不可诊断。
临床应用与阅片要点
- 在对比剂减量50%的静脉期胸部双层探测器CT中,可重建40 keV单能图像以改善肺门淋巴结 delineation 与整体对比增强。
- 作者认为减量对比剂方案可考虑用于更广泛的双层探测器CT检查与患者。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Intravenous contrast administration is crucial in many CT examinations but also poses a potential risk to the patient. Monoenergetic images (MonoE) of dual-energy CT systems can virtually increase iodine attenuation and might improve image quality (IQ) if contrast dose is reduced. In this study, we investigated the influence of MonoE on lymph node (LN) delineation and IQ in chest CT examinations with significantly reduced contrast dose (50 %) of a novel dual-layer CT (DLCT).
METHOD: 30 patients with clinically indicated reduced contrast dose underwent venous-phase chest DLCT scans. Conventional polyenergetic (PolyE) and MonoE images at 40 keV were calculated. The contrast difference of hilar lymph nodes (LN-CD) to the adjacent right pulmonary artery, their signal-to-noise (SNR) and contrast-to-noise-ratio (CNR) were determined. Subjective IQ was evaluated by 2 readers with respect to LN delineation and overall contrast enhancement (CE) using a 5-point-Likert-scale.
RESULTS: LN-CD, SNR and CNR were significantly higher in MonoE than in PolyE images (LN-CD 92.3 ± 37.9 vs. 33.1 ± 14.5 HU, SNR 8.4 ± 3.4 vs. 4.0 ± 1.2, CNR 9.2 ± 6.3 vs. 2.6 ± 1.5; all p < 0.01). The LN delineation (3.7 ± 0.9 vs.1.8 ± 0.7; p < 0.01) and the CE (3.9 ± 0.7 vs. 2.3 ± 0.7; p < 0.01) were rated significantly better for MonoE than for PolyE images. There was no MonoE examination classified as non-diagnostic.
CONCLUSIONS: Subjective and objective IQ parameters can be significantly improved for venous-phase chest CT examinations with reduced contrast doses by utilization of low-keV MonoE reconstructions. All MonoE images provided sufficient overall CE and therefore reduced contrast doses might be considered in a wider range of DLCT examinations and patients.