新型无线供电脊髓电刺激导线及接收器的MRI安全性评估

Evaluation of MRI Issues for a New Wirelessly Powered, Spinal Cord Stimulation Lead With Receiver.
PubMed 文献 · AJR Am J Roentgenol · 2019-12-17

该体外研究评估了用于脊髓电刺激的新型无线供电导线及接收器的MRI安全性,包括磁场相互作用、射频致热、功能干扰和伪影。3 T下导线呈轻度磁场相互作用(偏转22°、无扭矩),1.5 T/64 MHz与3 T/128 MHz下最高温升分别为3.8°C和11.3°C,未造成功能损害,梯度回波序列最大伪影约为导线尺寸外延10 mm。作者结论为在特定全身平均SAR限值(1.5 T/64 MHz下2.0 W/kg、3 T/128 MHz下0.3 W/kg)条件下,植入该系统的患者可安全接受MRI检查。

要点

  • 3 T下导线偏转角度22°、无扭矩,属轻度磁场相互作用。
  • 1.5 T/64 MHz与3 T/128 MHz下最高温升分别为3.8°C与11.3°C。
  • 梯度回波序列最大伪影约超出导线尺寸10 mm。

核心结论

  • 作者结论为在该研究提供的安全指南条件下,该新型无线供电导线及接收器未发现实质性担忧。
  • 在适当全身平均SAR水平(1.5 T/64 MHz下2.0 W/kg、3 T/128 MHz下0.3 W/kg,对应调整后温升3.6°C与1.2°C)下,MRI检查致热可接受。
  • 作者认为植入该无线供电导线及接收器的患者可在特定条件下安全接受MRI检查。

研究设计与样本

  • 体外研究,对用于脊髓电刺激的无线供电导线及接收器样品采用标准化技术进行MRI安全性评估。
  • 磁场相互作用(平移吸引力和扭矩)与伪影在3 T下测试。
  • MRI相关致热在1.5 T/64 MHz和3 T/128 MHz下采用数值模拟与分析建模以及物理测试两种方法进行。
  • 功能干扰在1.5 T/64 MHz和3 T/128 MHz的MRI暴露条件下评估。

关键影像表现/征象

  • 3 T下导线表现为轻度磁场相互作用,偏转角度22°,无扭矩。
  • 1.5 T/64 MHz下记录到的最高温度变化为3.8°C。
  • 3 T/128 MHz下记录到的最高温度变化为11.3°C。
  • 暴露于MRI条件未损坏或改变导线的功能特性。
  • 梯度回波脉冲序列上见到的最大伪影约超出导线尺寸10 mm。

诊断标准与关键数值

  • 安全指南基于本研究结果给出:1.5 T/64 MHz下全身平均SAR为2.0 W/kg,对应调整后温升3.6°C。
  • 安全指南基于本研究结果给出:3 T/128 MHz下全身平均SAR为0.3 W/kg,对应调整后温升1.2°C。

临床应用与阅片要点

  • 作者结论认为,植入该无线供电导线及接收器的患者可在特定条件下安全接受MRI检查。
  • 作者给出的特定条件为全身平均SAR在1.5 T/64 MHz下不超过2.0 W/kg、在3 T/128 MHz下不超过0.3 W/kg。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. MRI is an imaging modality frequently ordered for patients with neuromodulation systems implanted for spinal cord stimulation. The purpose of this investigation was to evaluate MRI safety issues (magnetic field interactions, MRI-related heating, functional disturbances, and artifacts) for a new wirelessly powered lead with receiver used for SCS. MATERIALS AND METHODS. Lead samples underwent in vitro evaluation for MRI safety issues using standardized techniques. Magnetic field interactions (i.e., translational attraction and torque) and artifacts were tested at 3 T. MRI-related heating was performed at 1.5 T/64 MHz and 3 T/128 MHz using two different methods: numerical simulations with analytical modeling and physical testing. Possible functional disturbances were evaluated under exposures to 1.5-T/64-MHz and 3-T/128-MHz MRI conditions. RESULTS. The lead exhibited minor magnetic field interactions (22° deflection angle, no torque) at 3 T. The highest temperature change recorded at 1.5 T/64 MHz and 3 T/128 MHz was 3.8°C and 11.3°C, respectively. Exposures to MRI conditions did not damage or alter the functional aspects of the leads. The maximum artifact size seen on a gradient-echo pulse sequence extended approximately 10 mm relative to the size of the lead. CONCLUSION. The MRI tests performed on patients with the new lead with receiver revealed no substantial concerns with respect to the conditions that we provide in the safety guidelines that were based on the results of this investigation. Therefore, MRI examinations will result in acceptable heating when conducted at appropriate whole-body-averaged specific absorption rate levels (i.e., 2.0 W/kg at 1.5 T/64 MHz and 0.3 W/kg at 3 T/128 MHz, corresponding to adjusted temperature rises of 3.6°C and 1.2°C, respectively). Therefore, patients with this wirelessly powered lead and receiver implanted can safely undergo MRI examinations under specific conditions.