当代透视引导介入操作的患者有效剂量评估

Effective Dose Assessment for Patients Undergoing Contemporary Fluoroscopically Guided Interventional Procedures.
PubMed 文献 · AJR Am J Roentgenol · 2019-10-31

该回顾性研究收集了2016年5月至2018年10月12间检查室连续成人病例的剂量数据,共24,911例,按操作分类,利用KAP和操作特异性转换系数计算有效剂量。结果给出了101种操作的KAP和有效剂量,其中12种常见操作的中位KAP<3.26 Gy·cm2、中位有效剂量<0.70 mSv,但部分少见操作剂量较高,如门静脉造影的KAP第95百分位达985 Gy·cm2、有效剂量第95百分位达256 mSv。作者认为这些数值有助于临床研究方案设计、剂量管理及在缺乏国家基准时评估合规性。

要点

  • 常见介入操作(占病例≥50%的12种)中位KAP<3.26 Gy·cm2,中位有效剂量<0.70 mSv。
  • 部分少见操作剂量较高,如门静脉造影KAP第95百分位≥985 Gy·cm2、有效剂量第95百分位256 mSv。
  • NCRP报告168和172未提供89种操作的KAP基准及全部101种操作的有效剂量基准。

诊断标准与关键数值

  • 12种占病例至少50%的操作中位KAP<3.26 Gy·cm2,中位有效剂量<0.70 mSv。
  • 16种操作KAP第95百分位≥500 Gy·cm2,门静脉造影KAP第95百分位为985 Gy·cm2。
  • 21种操作有效剂量第95百分位≥100 mSv,门静脉造影有效剂量第95百分位为256 mSv。

临床应用与阅片要点

  • 可利用本文提供的操作特异性KAP和有效剂量数值辅助设计临床研究方案和剂量管理项目。
  • 在缺乏国家基准的情况下,可参考本文数值评估机构对联合委员会透视服务标准的合规性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. The purpose of this study was to establish procedure-specific air kerma-area product (KAP) and effective dose for a large number of fluoroscopically guided interventional (FGI) procedures. MATERIALS AND METHODS. This retrospective study collected dose data for consecutive adult cases from 12 examination rooms between May 2016 and October 2018. A total of 24,911 cases (50.9% men) were categorized by procedure. Effective dose was calculated using KAP and procedure-specific KAP to effective dose conversion coefficients, mostly from National Council on Radiation Protection and Measurements (NCRP) Report 160. Data analysis was conducted with statistical software to determine mean value and five percentiles (10th, 25th, 50th, 75th, 95th). RESULTS. KAP and effective dose were presented for 101 procedures; a national benchmark is not available from NCRP Reports 168 and 172 for the KAP value of 89 procedures and for the effective dose of all 101 procedures. Twelve procedures that comprised at least 50% of patient cases had median KAP values less than 3.26 Gy · cm2 and a median effective dose of less than 0.70 mSv. However, some infrequent procedures might be associated with a higher dose. The 95th percentile of KAP was greater than or equal to 500 Gy · cm2 for 16 procedures and 985 Gy · cm2 for portography; for effective dose it was greater than or equal to 100 mSv for 21 procedures and 256 mSv for portography. CONCLUSION. The values for KAP and effective dose provided in this article can aid in design and review of clinical research protocols and dose management programs and in assessing compliance with the Joint Commission's standards for organizations providing fluoroscopy services in the absence of national benchmarks for more than 89 procedures.