文献速递神经介入CT

金属伪影减少算法改善支架辅助弹簧圈栓塞治疗的中大型脑动脉瘤锥形束CT图像质量

Metal artifact reduction algorithm for image quality improvement of cone-beam CT images of medium or large cerebral aneurysms treated with stent-assisted coil embolization.
PubMed 文献 · Neuroradiology · 2019-11-07

回顾性评估18例支架辅助弹簧圈栓塞治疗的中大型脑动脉瘤患者的锥形束CT数据,比较有无金属伪影减少(MAR)处理的主观与客观图像质量。MAR处理使17/18例(94.4%)四项主观标准的中位评分均改善,金属伪影影响区域的感兴趣区平均CT值显著降低。弹簧圈团最大直径<17 mm(OR 4.0,95%CI 1.2-13.9,p=0.02)和金属伪影覆盖血管长度<24 mm(OR 2.3,95%CI 1.1-4.7,p=0.03)与MAR的 usefulness 显著相关。

要点

  • MAR处理可改善支架辅助弹簧圈栓塞术后中大型动脉瘤锥形束CT的主观图像质量,17/18例(94.4%)四项标准中位评分改善。
  • MAR处理后金属伪影影响区域的感兴趣区平均CT值显著降低。
  • 弹簧圈团最大直径<17 mm和金属伪影覆盖血管长度<24 mm与MAR有用性显著相关。

关键影像表现/征象

  • MAR处理使17/18例(94.4%)四项主观标准的中位评分均改善。
  • 金属伪影影响区域的感兴趣区平均CT值在MAR处理后显著降低。

诊断标准与关键数值

  • 弹簧圈团最大直径<17 mm与MAR有用性显著相关(OR 4.0,95%CI 1.2-13.9,p=0.02)。
  • 金属伪影覆盖血管长度<24 mm与MAR有用性显著相关(OR 2.3,95%CI 1.1-4.7,p=0.03)。

临床应用与阅片要点

  • 作者认为MAR算法对于中大型动脉瘤支架辅助弹簧圈栓塞术后的锥形束CT图像质量改善具有可行性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The aim of the present study was to assess image quality improvement using a metal artifact reduction (MAR) algorithm in cases of medium or large cerebral aneurysms treated with stent-assisted coil embolization (SAC), and to analyze factors associated with the usefulness of the MAR algorithm.

METHODS: We retrospectively evaluated the cone-beam computed tomography (CBCT) data sets of 18 patients with cerebral aneurysms treated with SAC. For subjective analysis, images of all cases with and without MAR processing were evaluated by five neurosurgeons based on four criteria using a five-point scale. For objective analysis, the CT values of all cases with and without MAR processing were calculated. In addition, we assessed factors associated with the usefulness of the MAR by analyzing the nine cases in which the median score for criterion 1 improved by more than two points.

RESULTS: MAR processing improved the median scores for all four criteria in 17/18 cases (94.4%). Mean CT values of the region of interest at the site influenced by metal artifacts were significantly reduced after MAR processing. The maximum diameter of the coil mass (< 17 mm; odds ratio [OR], 4.0; 95% confidence interval [CI], 1.2-13.9; p = 0.02) and vessel length covered by metal artifacts (< 24 mm; OR, 2.3; 95% CI, 1.1-4.7; p = 0.03) was significantly associated with the usefulness of the MAR.

CONCLUSIONS: This study suggests the feasibility of a MAR algorithm to improve the image quality of CBCT images in patients who have undergone SAC for medium or large aneurysms.

来源:Neuroradiology,查看原文。版权归原作者所有。

引用

  1. Murai S, Hiramatsu M, Takasugi Y, et al. Metal artifact reduction algorithm for image quality improvement of cone-beam CT images of medium or large cerebral aneurysms treated with stent-assisted coil embolization. Neuroradiology. 2019. PMID: 31701181. doi:10.1007/s00234-019-02297-8

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。