超声造影定量肌肉灌注的观察者内与设备依赖的观察者间信度
研究在42例肩部进行166次超声造影定量三角肌灌注,观察者内ICC为0.91、CV为10.28%,观察者间ICC为0.84,使用不同超声设备时ICC降至0.60;高切面一致性亚组信度显著提高。作者认为超声造影灌注定量似乎可用于多中心研究,但不同设备合并分析仍存关键问题,切面一致性对信度和可迁移性似乎至关重要。
要点
- 观察者内信度高(ICC=0.91,CV=10.28%),观察者间信度ICC=0.84但不同设备时降至0.60。
- 切面一致性高的亚组信度显著提高,观察者内ICC=0.97、CV=5.49%。
- 不同超声设备间数据合并仍是多中心应用的关键限制。
核心结论
- 三角肌超声造影灌注定量似乎适用于多中心研究,但合并不同超声设备的数据仍存关键问题。
- 切面一致性似乎对超声造影肌肉灌注定量的信度和可迁移性至关重要。
研究设计与样本
- 在42例肩部进行166次超声造影定量,评估观察者内和观察者间信度,包括不同超声设备之间的可迁移性。
- 原文未报告中心数、具体设备型号及参考标准。
诊断标准与关键数值
- 观察者内信度:ICC=0.91,CV=10.28%。
- 观察者间信度:ICC=0.84,CV=17.1%;使用不同超声设备时ICC=0.60,CV=18.6%。
- 高切面一致性亚组:观察者内ICC=0.97、CV=5.49%;观察者间同设备ICC=0.98、CV=5.83%;不同设备ICC=0.78、CV=9.8%。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Muscle perfusion quantification by contrast-enhanced ultrasound (CEUS) may facilitate treatment decisions in musculoskeletal disorders. Translation into clinical routine relies on high intra-observer and inter-observer reliability and transferability between ultrasound devices to enable validation and multicenter studies. This study evaluates these aspects for deltoid muscle perfusion quantification, including possible multicenter study setups. One hundred sixty-six CEUS quantifications were conducted on 42 shoulders. Intra-observer reliability revealed a high intra-class correlation coefficient (ICC, r = 0.91) and low coefficient of variation (CV, 10.28%). Inter-observer reliability revealed an ICC of .84 and a CV of 17.1%, but these values decreased when different ultrasound devices were used (ICC = .60, CV = 18.6%). Re-evaluating subgroups with high sectional plane concordance significantly increased reliability (intra-observer: ICC = .97, CV = 5.49%, inter-observer/same device: ICC = .98, CV = 5.83%, varying devices: ICC = .78, CV = 9.8%). CEUS perfusion quantification of the deltoid seems applicable for multicenter studies, yet pooling different ultrasound devices remains critical. Sectional plane concordance appears to be crucial for reliability and transferability of CEUS muscle perfusion quantifications.
引用
- Kunz P, Kiesl S, Groß S, et al. Intra-observer and Device-Dependent Inter-observer Reliability of Contrast-Enhanced Ultrasound for Muscle Perfusion Quantification. Ultrasound Med Biol. 2019. PMID: 31733932. doi:10.1016/j.ultrasmedbio.2019.10.007
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。