文献速递腹部介入技术与安全MR

基于患者数据的肋骨三维模型用于高强度聚焦超声经肋骨治疗:初步研究

Patient-Specific 3-Dimensional Model for High-Intensity Focused Ultrasound Treatment Through the Rib Cage: A Preliminary Study.
PubMed 文献 · J Ultrasound Med · 2019-11-13

本研究利用患者数据重建肋骨三维模型,结合组织仿体与磁共振引导高强度聚焦超声多阵元相控换能器进行实验,评估肋骨对声压分布和声功率的影响。结果显示肋骨对声场的影响与肋骨解剖结构密切相关,焦点温升变化与肋骨模型至焦点的距离及肋骨在声束路径上遮挡换能器元件的阴影面积有关。

要点

  • 肋骨对声压分布和声功率的影响与肋骨解剖结构密切相关。
  • 焦点温升变化与肋骨模型至焦点距离及肋骨遮挡换能器元件的阴影面积有关。
  • 可通过改变HIFU功率或作用时间来调控焦点加热。

核心结论

  • 实验结果表明肋骨模型是有价值的工具,可提供逼真的人体解剖结构和属性以评估肋骨对超声传播的影响。

研究设计与样本

  • 初步实验研究,利用基于患者数据重建的肋骨模型、组织仿体及磁共振引导高强度聚焦超声多阵元相控换能器进行实验。
  • 使用ExAblate系统(InSightec Ltd,以色列Tirat Carmel)对仿体进行声处理,改变目标仿体或肋骨模型的位置。
  • 实验旨在推导用于定义上腹部肋骨下疾病治疗协议中所需能量和调整量的参数。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The purpose of this study was to develop a patient-specific 3-dimensional model for high-intensity focused ultrasound (HIFU) treatment through the rib cage using patient data.

METHODS: Experimental testing to derive parameters used in defining the amount of energy and alteration needed in treatment protocols for upper abdominal disorders under the rib cage was performed. Reconstructed rib cage models based on patient data, tissue-mimicking material phantoms, and magnetic resonance imaging-guided HIFU using a multielement phased array transducer were used in the experiments. Changes in the focal temperature, acoustic power, and acoustic pressure distribution were investigated with and without the presence of the rib cage model. An ExAblate system (InSightec Ltd, Tirat Carmel, Israel) was used to sonicate phantoms by varying the target phantom or rib cage model location.

RESULTS: The effect of the rib cage on the acoustic pressure distribution and acoustic power was closely related to the anatomic structures of the ribs. Thermometry revealed that heating at the focus could be controlled by changing either the power or duration of HIFU application to improve the focal temperature change. The focal temperature change was found to be related to the distance between the rib cage model and focus and the shadow area on the transducer elements covered by the rib cage model in the beam path.

CONCLUSIONS: Experimental results suggest that the rib cage model is a valuable and useful tool that can provide realistic human anatomic structures and properties for evaluating the effects of the rib cage on ultrasound propagation.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Cao R, Huang Z, Nabi G, et al. Patient-Specific 3-Dimensional Model for High-Intensity Focused Ultrasound Treatment Through the Rib Cage: A Preliminary Study. J Ultrasound Med. 2019. PMID: 31721248. doi:10.1002/jum.15170

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。