肝脏MR弹性成像在不同场强、脉冲序列、扫描间隔和阅片者间的可重复性
前瞻性纳入成年非酒精性脂肪性肝病患者,同日及2周后分别行1.5 T 2D GRE和3.0 T 2D SE-EPI肝脏MRE,每位患者共4次检查,由3名不同背景阅片者测量肝硬度。汇总可重复性系数(RDC)为26%(95%CI上限30.6%),一致性界限在0.55 kPa以内,不同场强/序列在2周间隔下可重复性较低。
要点
- 肝脏MRE肝硬度测量在跨场强、序列和2周间隔时汇总RDC为26%(95%CI上限30.6%)。
- 不同场强/序列在较长(2周)间隔下测量可重复性较低。
- 阅片者间平均偏差小(0.05-0.19 kPa),一致性界限窄(±0.53 kPa)。
核心结论
- 跨扫描参数组合的变化幅度在临床可接受范围内,肝脏MRE肝硬度可作为可重复的定量影像生物标志物。
- 在不同场强/MRE序列且间隔较长(2周)时,测量可重复性较低。
- 操作者应接受培训,在重复检查中一致地勾画感兴趣区。
研究设计与样本
- 前瞻性研究,纳入完成MRE检查的成年非酒精性脂肪性肝病患者20例。
- 每位患者共行4次MRE:同日及2周后分别行1.5 T 2D GRE序列和3.0 T 2D SE-EPI序列。
- 由3名对MRE技术和肝脏解剖背景知识水平不同的阅片者在培训后测量肝硬度。
诊断标准与关键数值
- 跨场强、MRE序列和2周扫描间隔、3名阅片者的汇总RDC为26%(95%CI上限30.6%),一致性界限在0.55 kPa以内。
- 阅片者间平均偏差为0.05-0.19 kPa ± 0.53。
临床应用与阅片要点
- 操作者应接受培训,在重复检查中一致地勾画感兴趣区。
局限性
- 在不同场强/MRE序列且间隔较长(2周)时,测量可重复性较低。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate the reproducibility of hepatic MRE under various combinations of settings of field strength, pulse sequence, scan interval, and reader in non-alcoholic fatty liver disease (NAFLD) patients.
METHODS: Adult NAFLD patients were prospectively enrolled for serial hepatic MRE with 1.5 T using 2D GRE sequence and 3.0 T using 2D SE-EPI sequence on the same day and after 2 weeks, resulting a total of four MRE examinations per patient. Three readers with various levels of background knowledge in MRE technique and liver anatomy measured liver stiffness after a training session. Linear regression, Bland-Altman analysis, within-subject coefficient of variation, and reproducibility coefficient (RDC) were used to determine reproducibility of hepatic MRE measurement.
RESULTS: Twenty patients completed the MRE sessions. Liver stiffness through MRE showed pooled RDC of 26% (upper 95% CI 30.6%) and corresponding limits of agreement (LOA) within 0.55 kPa across field strengths, MRE sequences, and 2-week interscan interval in three readers. Small mean biases and narrow LOA were observed among readers (0.05-0.19 kPa ± 0.53).
CONCLUSION: The magnitude of change across combinations of scan parameters is within acceptable clinical range, rendering liver stiffness through MRE a reproducible quantitative imaging biomarker. A lower reproducibility was observed for measurements under different field strengths/MRE sequences at a longer (2 weeks) interscan interval. Operators should be trained to acquire region of interest consistently in repeat examinations.