三叶式与二叶式主动脉瓣狭窄患者接受生物瓣与机械瓣置换术后升主动脉的向量血流成像
本研究对10例三叶式主动脉瓣狭窄拟行生物瓣置换、10例二叶式主动脉瓣狭窄拟行机械瓣置换及10例正常主动脉瓣患者,术中于升主动脉行向量血流成像(VFI)与常规频谱多普勒检查。二叶式狭窄血流较三叶式更复杂(p<0.02),两组均较正常瓣膜血流更复杂且更快(p<0.0001);术后机械瓣与生物瓣血流参数均较术前改变(p<0.0001),机械瓣旋转频率较生物瓣慢(p<0.01),所有植入物均未恢复至正常血流(p<0.0001)。
要点
- VFI可无角度依赖地量化血流复杂度(VC)与次级旋转(SR),为AS及瓣膜植入物评估提供新参数。
- 二叶式AS较三叶式AS升主动脉血流更复杂(p<0.02)。
- 术后机械瓣SR较生物瓣慢(p<0.01),所有植入物均未达正常血流(p<0.0001)。
关键影像表现/征象
- 二叶式AS(PSV 380.9 cm/s,SR 16.7 Hz,VC 0.21)血流较三叶式AS(PSV 346.1 cm/s,SR 17.1 Hz,VC 0.33)更复杂(p<0.02)。
- 三叶式与二叶式AS两组血流均较正常主动脉瓣(PSV 124.0 cm/s,SR 4.3 Hz,VC 0.79)更复杂且更快(p<0.0001)。
- 术后机械瓣(PSV 250.4 cm/s,SR 9.8 Hz,VC 0.54)与生物瓣(PSV 232.4 cm/s,SR 12.5 Hz,VC 0.61)血流参数均较术前改变(p<0.0001)。
- 机械瓣SR较生物瓣慢(p<0.01)。
- 所有植入物均未恢复至正常血流(p<0.0001)。
诊断标准与关键数值
- VC与PSV相关(r=0.87),SR与PSV相关(r=0.89)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Aortic valve stenosis (AS) is treated with biological prostheses (BPs) and mechanical prostheses (MPs). Vector flow imaging (VFI), an angle-independent ultrasound method, can quantify flow complexity (vector concentration (VC)) and secondary rotation (SR). Ten patients (mean age: 70.7 y) with tricuspid AS scheduled for BPs, 10 patients (mean age: 56.2 y) with bicuspid AS scheduled for MPs and 10 patients (mean age: 63.9 y) with normal aortic valves were scanned intra-operatively on the ascending aorta with VFI and conventional spectral Doppler. Bicuspid AS (peak systolic velocity (PSV): 380.9 cm/s, SR: 16.7 Hz, VC: 0.21) had more complex flow (p < 0.02) than tricuspid AS (PSV: 346.1 cm/s, SR: 17.1 Hz, VC: 0.33). Both groups had more complex and faster flow (p < 0.0001) than normal aortic valve patients (PSV: 124.0 cm/s, SR: 4.3 Hz, VC: 0.79). VC (r = 0.87) and SR (r = 0.89) correlated to PSV. After surgery, flow parameters changed (p < 0.0001) for patients with MPs (PSV: 250.4 cm/s, SR: 9.8 Hz, VC: 0.54) and BPs (PSV: 232.4 cm/s, SR: 12.5 Hz, VC: 0.61), with MPs having slower SR (p < 0.01). None of the implants had normal flow (p < 0.0001). In conclusion, VFI can provide new flow parameters for AS and implant assessment.
引用
- Hansen KL, Møller-Sørensen H, Kjaergaard J, et al. Vector Flow Imaging of the Ascending Aorta in Patients with Tricuspid and Bicuspid Aortic Valve Stenosis Treated with Biological and Mechanical Implants. Ultrasound Med Biol. 2019. PMID: 31677849. doi:10.1016/j.ultrasmedbio.2019.09.020
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。