第三代双源CT超高螺距协议用于TAVI规划与冠状动脉树评估:可行性、图像质量与诊断效能

Third generation dual source CT with ultra-high pitch protocol for TAVI planning and coronary tree assessment: feasibility, image quality and diagnostic performance.
PubMed 文献 · Eur J Radiol · 2019-11-13

前瞻性研究纳入223例接受DSCT的TAVI患者,采用单次前瞻性心电触发超高螺距采集,评估图像质量并以有创冠状动脉造影为参考标准评价冠状动脉狭窄诊断效能。结果显示各节段衰减值均>400HU,图像质量中位评分≥3,狭窄≥50%时基于患者的敏感度97.8%、特异度88.8%,狭窄≥70%时敏感度92.5%、特异度85.8%;钙化负荷较低组CT表现更好,主动脉瓣环CT面积与植入假体尺寸直接相关。

要点

  • 单次前瞻性心电触发超高螺距DSCT采集用于TAVI规划可行,各节段衰减值均>400HU且图像质量中位评分≥3。
  • 以ICA为参考,狭窄≥50%时基于患者的敏感度97.8%、特异度88.8%、准确率90.6%。
  • 严重冠状动脉钙化患者CT诊断表现较差,阅片时需警惕钙化所致假阳性。

核心结论

  • 第三代192×2双源CT采用单次前瞻性心电触发超高螺距采集,用于TAVI规划及冠状动脉狭窄评估是可行的。
  • CT在冠状动脉钙化评分较低的患者中表现更好,在严重冠状动脉钙化患者中表现较差。
  • 主动脉瓣环CT衍生面积与植入假体尺寸之间存在直接相关。

研究设计与样本

  • 前瞻性研究,纳入223例接受DSCT检查的TAVI患者。
  • 采用第三代192×2双源CT,单次前瞻性心电触发超高螺距采集协议。
  • 计算冠状动脉钙化评分,测量主动脉各水平、股动脉及冠状动脉的衰减值,图像质量采用4分制评估。
  • 以有创冠状动脉造影为参考标准,比较CT对冠状动脉狭窄≥50%和≥70%的评估效能。
  • 主动脉瓣环CT衍生面积与植入假体尺寸的相关性采用Spearman检验分析。

关键影像表现/征象

  • 所有节段获得的衰减值均>400HU。
  • 图像质量中位评分≥3(4分制)。

诊断标准与关键数值

  • 狭窄≥50%节段分析:敏感度97.6%、特异度87.6%、阳性预测值64.2%、阴性预测值99.0%、诊断准确率89.6%。
  • 狭窄≥50%基于患者分析:敏感度97.8%、特异度88.8%、阳性预测值68.8%、阴性预测值99.4%、诊断准确率90.6%。
  • 狭窄≥70%节段分析:敏感度88.5%、特异度83.8%、阳性预测值54.7%、阴性预测值96.8%、诊断准确率84.8%。
  • 狭窄≥70%基于患者分析:敏感度92.5%、特异度85.8%、阳性预测值58.7%、阴性预测值98.1%、诊断准确率87.0%。
  • CT在冠状动脉钙化评分较低组中表现更好;主动脉瓣环CT衍生面积与植入假体尺寸呈直接相关(Spearman检验)。

临床应用与阅片要点

  • 单次前瞻性心电触发超高螺距采集可用于TAVI术前规划中的主动脉瓣环测量及冠状动脉树评估。
  • CT对冠状动脉狭窄≥50%具有较高的阴性预测值(基于患者99.4%),可用于排除显著狭窄。
  • 严重冠状动脉钙化患者中CT诊断表现较差,解读时需考虑钙化对狭窄评估的影响。
  • 主动脉瓣环CT衍生面积与假体尺寸直接相关,可供TAVI术前假体选择参考。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the feasibility, image quality (IQ) and diagnostic performance of third generation 192 × 2 dual source computer tomography (DSCT) with ultra-high pitch acquisition for trans-catheter aortic valve implantation (TAVI) planning and coronary tree assessment.

METHOD: In this prospective study, 223 patients underwent to DSCT for TAVI. Coronary calcium scoring (CCS) was calculated. Attenuation values were measured at aortic levels, femoral and coronary arteries. IQ was evaluate with a 4-point scale. The CT performance, in the assessment of coronary stenosis ≥50 % and ≥70 %, was compared with invasive coronary angiography (ICA), served as reference standard. Aortic annulus (AoA) CT derived area and implanted prosthesis size were correlate with Spearman's test.

RESULTS: Attenuation values >400HU were obtain in all segments. IQ median value was ≥ 3. In the assessment of stenosis ≥50 %, on a segment-based analysis, CT sensitivity, specificity, positive and negative predictive values and diagnostic accuracy were 97.6 %, 87.6 %, 64.2 %, 99.0 % and 89.6 %, on patient-based analysis were 97.8 %, 88.8 %, 68.8 %, 99.4 % and 90.6 %, respectively. In the assessment of stenosis ≥70 %, on segment-based analysis, were 88.5 %, 83.8 %, 54.7 %, 96.8 % and 84.8 %, and on patient-based analysis were 92.5 %, 85.8 %, 58.7 %, 98.1 % and 87.0 %, respectively. The CT performed better in the group with lower CCS. A direct correlation was found between AoA CT derived area and prosthesis size.

CONCLUSION: DSCT, using a single prospective ECG-triggered ultra-high pitch acquisition, is feasible for TAVI planning and in the assessment of coronary stenosis. CT performed worse in patients with severe coronary calcifications.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Schicchi N, Fogante M, Pirani PE, et al. Third generation dual source CT with ultra-high pitch protocol for TAVI planning and coronary tree assessment: feasibility, image quality and diagnostic performance. Eur J Radiol. 2019. PMID: 31759224. doi:10.1016/j.ejrad.2019.108749

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。