文献速递心血管胸部CT

Turbo-flash与回顾性双源CT冠状动脉造影的活体辐射剂量学与图像质量比较

In vivo radiation dosimetry and image quality of turbo-flash and retrospective dual-source CT coronary angiography.
PubMed 文献 · Radiol Med · 2019-11-05

前瞻性纳入68例患者行192×2双源CT冠状动脉成像,31例采用turbo-flash方案、37例采用回顾性方案,用热释光剂量计测量活体辐射剂量并比较图像质量。turbo-flash方案在甲状腺、心脏、左乳晕及左半胸的测量剂量及CTDIvol、SSDE、DLP、有效剂量均显著低于回顾性方案,图像质量差异无统计学意义。

要点

  • turbo-flash方案心脏水平测量剂量3.58 mGy,显著低于回顾性方案的28.72 mGy。
  • 两方案图像质量差异无统计学意义,观察者间一致性分别为良好(k=0.796)与很好(k=0.817)。
  • 各方案中测量剂量与SSDE均呈极好相关(turbo-flash r=0.9298,回顾性 r=0.9753)。

核心结论

  • turbo-flash方案的测量剂量、CTDIvol、SSDE、DLP和有效剂量均显著低于回顾性方案,图像质量相似,测量剂量与SSDE在每种方案中均呈极好相关。

研究设计与样本

  • 前瞻性研究,共纳入68例患者(平均年龄59.2±9.7岁),使用192×2双源CT(SOMATOM Force,Siemens)排除冠状动脉疾病。
  • 31例采用turbo-flash方案,37例采用回顾性方案。
  • 活体测量剂量通过将热释光剂量计置于甲状腺水平、心脏水平、左乳晕及左半胸浅表处获得。
  • 采用t检验比较两方案的各部位测量剂量及CTDIvol、SSDE、DLP、有效剂量等估算剂量参数,并比较图像质量。
  • 观察者间一致性采用kappa系数(k)评估,各方案中测量剂量与SSDE的相关性采用Pearson系数(r)评估。

诊断标准与关键数值

  • turbo-flash方案与回顾性方案的测量剂量:甲状腺1.43 vs. 2.58 mGy(p=0.0408),心脏3.58 vs. 28.72 mGy(p<0.0001),左乳晕3.00 vs. 24.21 mGy(p<0.0001),左半胸2.68 vs. 24.03 mGy(p<0.0001),turbo-flash方案均显著降低。
  • CTDIvol、SSDE、DLP和有效剂量在turbo-flash方案均显著低于回顾性方案。
  • 两方案图像质量差异无统计学意义。
  • 观察者间一致性:turbo-flash方案良好(k=0.796),回顾性方案很好(k=0.817)。
  • 测量剂量与SSDE相关性:turbo-flash方案r=0.9298(p<0.0001),回顾性方案r=0.9753(p<0.0001)。

临床应用与阅片要点

  • 在行冠状动脉CT成像时,turbo-flash方案可显著降低甲状腺、心脏、左乳晕及左半胸的测量剂量以及CTDIvol、SSDE、DLP和有效剂量,而图像质量与回顾性方案相似。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare measured radiation dose (MD), estimated radiation dose (ED) and image quality in coronary computed tomography between turbo-flash (TFP) and retrospective protocol (RP) and correlate MD with size-specific dose estimates (SSDE).

MATERIALS AND METHODS: In this prospective study, we selected 68 patients (mean age, 59.2 ± 9.7 years) undergoing 192 × 2 dual-source CT (SOMATOM Force, Siemens) to rule out coronary artery disease. Thirty-one underwent TFP and 37 RP. To evaluate in vivo MD, thermoluminescent dosimeters were placed, superficially, at thyroid and heart level, left breast areola and left hemi-thorax. MD in each site, and ED parameters, such as volume CT dose index (CTDIvol), SSDE, dose length product (DLP), effective dose (E), were compared between two protocols with a t test. Image quality was compared between two protocols. Inter-observer agreement was evaluated with a kappa coefficient (k). In each protocol, MD was correlated with SSDE using a Pearson coefficient (r).

RESULTS: Comparing TFP and RP, MD at thyroid (1.43 vs. 2.58 mGy; p = 0.0408), heart (3.58 vs. 28.72 mGy; p < 0.0001), left breast areola (3.00 vs. 24.21 mGy; p < 0.0001) and left hemi-thorax (2.68 vs. 24.03 mGy; p < 0.0001), CTDIvol, SSDE, DLP and E were significantly lower. Differences in image quality were not statistically significant. Inter-observer agreement was good (k = 0.796) in TFP and very good (k = 0.817) in RP. MD and SSDE excellently correlated with TFP (r = 0.9298, p < 0.0001) and RP (r = 0.9753, p < 0.0001).

CONCLUSIONS: With TFP, MD, CTDIvol, SSDE, DLP and E were significantly lower, than with RP. Image quality was similar between two protocols. MD correlated excellently with SSDE in each protocol.

来源:Radiol Med,查看原文。版权归原作者所有。

引用

  1. Schicchi N, Mari A, Fogante M, et al. In vivo radiation dosimetry and image quality of turbo-flash and retrospective dual-source CT coronary angiography. Radiol Med. 2019. PMID: 31686317. doi:10.1007/s11547-019-01103-y

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。