后处理降噪在CT肺气肿密度测量中的剂量降低潜力:体模研究
本研究用COPDGene体模在第三代双源CT上以16种扫描方案(CTDIvol 0.035–10.680 mGy)扫描,分析384个重建图像。结果显示降低辐射剂量使CT密度直方图重叠呈指数增加,软核重建、迭代重建和降噪软件均可减少重叠;中等级迭代重建联合降噪软件可在维持可接受图像质量的同时降低85%辐射剂量。
要点
- CT密度直方图重叠度可量化肺气肿与正常肺组织的可区分程度。
- 软核重建比硬滤波核产生更少的直方图重叠。
- 中等级迭代重建联合降噪软件可降低85%辐射剂量而维持可接受图像质量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The aim of this phantom study was to investigate the effect of scan parameters and noise suppression techniques on the minimum radiation dose for acceptable image quality for CT emphysema densitometry.
METHODS: The COPDGene phantom was scanned on a third generation dual-source CT system with 16 scan setups (CTDIvol 0.035-10.680 mGy). Images were reconstructed at 1.0/0.7 mm slice thickness/increment, with three kernels (one soft, two hard), filtered backprojection and three grades of third-generation iterative reconstruction (IR). Additionally, deep learning-based noise suppression software was applied. Main outcomes: overlap in area of the normalized histograms of CT density for the emphysema insert and lung material, and the radiation dose required for a maximum of 4.3% overlap (defined as acceptable image quality).
RESULTS: In total, 384 scan reconstructions were analyzed. Decreasing radiation dose resulted in an exponential increase of the overlap in normalized histograms of CT density. The overlap was 11-91% for the lowest dose setting (CTDIvol 0.035mGy). The soft kernel reconstruction showed less histogram overlap than hard filter kernels. IR and noise suppression also reduced overlap. Using intermediate grade IR plus noise suppression software allowed for 85% radiation dose reduction while maintaining acceptable image quality.
CONCLUSION: CT density histogram overlap can quantify the degree of discernibility of emphysema and healthy lung tissue. Noise suppression software, IR, and soft reconstruction kernels substantially decrease the dose required for acceptable image quality.
ADVANCES IN KNOWLEDGE: Noise suppression software, IR, and soft reconstruction kernels allow radiation dose reduction by 85% while still allowing differentiation between emphysema and normal lung tissue.
引用
- Wisselink HJ, Pelgrim GJ, Rook M, et al. Potential for dose reduction in CT emphysema densitometry with post-scan noise reduction: a phantom study. Br J Radiol. 2019. PMID: 31724436. doi:10.1259/bjr.20181019