减少能量切换与治疗室切换时间对多室质子治疗中心吞吐效率影响的回顾性分析
本研究回顾性分析某四室质子治疗中心软件升级前后各4个月的数据,能量切换时间由约4秒降至约0.5秒、治疗室切换时间由约30秒降至约20秒,共分析16168和17102个照射野。结果显示头颈及前列腺患者的束流等待时间降低近3倍、束流配送时间降低近2.5倍,16小时运行下机架容量由约30例增至40-45例。
要点
- 能量切换时间由约4秒降至约0.5秒、治疗室切换时间由约30秒降至约20秒后,束流等待时间降低近3倍、束流配送时间降低近2.5倍。
- 16小时日运行下,机架容量从约30例患者增至40-45例患者。
- 进一步效率提升的更大潜力来自摆位与成像效率的改善。
核心结论
- 减少能量切换时间和治疗室切换时间可显著增加质子治疗中心容量,进一步增益的更大潜力来自摆位与成像效率的改善。
研究设计与样本
- 回顾性分析,某四室质子治疗中心每日治疗约90例患者、已优化运行约7年,比较软件升级前后各4个月、患者量相当的16168和17102个照射野。
临床应用与阅片要点
- 作者指出许多质子中心正致力于提高效率,本研究显示减少能量与治疗室切换时间可显著增加中心容量,进一步增益的更大潜力来自摆位与成像效率的改善。
局限性
- 原文为回顾性分析。
- 原文在结论中提出进一步增益的更大潜力来自摆位与成像效率的改善,未报告其他局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To quantify how a control software upgrade changed beam delivery times and impacted efficiency and capacity of a multiroom proton therapy center.
METHODS: A four-room center treating approximately 90 patients/day, treating for approximately 7 years with optimized operations, underwent a software upgrade which reduced room and energy switching times from approximately 30 to 20 s and approximately 4 s to ~0.5 s, respectively. The center uses radio-frequency identification data to track patient treatments and has software which links this to beam delivery data extracted from the treatment log server. Two 4-month periods, with comparable patient volume, representing periods before and after the software change, were retrospectively analyzed.
RESULTS: A total of 16,168 and 17,102 fields were analyzed. For bilateral head and neck and prostate patients, the beam waiting time was reduced by nearly a factor of 3 and the beam delivery times were reduced by nearly a factor of 2.5. Room switching times were reduced more modestly. Gantry capacity has increased from approximately 30 patients to 40-45 patients in a 16-h daily operation.
CONCLUSIONS: Many proton centers are striving for increased efficiencies. We demonstrated that reductions in energy and room switching time can significantly increase center capacity. Greater potential for further gains would come from improvements in setup and imaging efficiency.
ADVANCES IN KNOWLEDGE: This paper provides detailed measured data on the effect on treatment times resulting from reducing energy and room switching times under controlled conditions. It helps validate the models of previous investigations to establish treatment capacity of a proton therapy center.