文献速递头颈神经CT

附加锡滤过的新型光子计数探测器CT用于鼻窦与颞骨成像的辐射剂量降低

Dose Reduction for Sinus and Temporal Bone Imaging Using Photon-Counting Detector CT With an Additional Tin Filter.
PubMed 文献 · Invest Radiol · 2020-01-01

该研究在体模、尸头及鼻窦与颞骨患者中评估附加锡滤过(Sn-100 kV)的高分辨率光子计数探测器CT(PCD-CT)相比能量积分探测器CT(EID-CT)的辐射剂量降低效果,体模显示噪声降低26%对应剂量降低56%,患者鼻窦检查平均剂量降低67%,颞骨检查剂量降低最高达85%,且10与8 mGy的PCD-CT图像对关键鼻窦结构的显示显著优于临床EID-CT。

要点

  • Sn-100 kV PCD-CT在鼻窦患者中平均剂量降低67%,颞骨患者中最高降低85%。
  • 10与8 mGy的PCD-CT鼻窦图像对腭小孔、鼻上颌缝、筛前动脉管显示显著优于临床EID-CT。
  • 6与7 mGy的PCD-CT与EID-CT图像间无显著差异,内耳结构显示相似。

核心结论

  • 体模与尸头研究显示Sn-100 kV PCD-CT相比现行临床EID-CT可在保持图像对比度的同时降低辐射剂量,并在鼻窦与颞骨患者研究中得到验证。
  • PCD-CT的超高分辨率能力相比现行临床标准改善了鼻窦成像的解剖结构显示。

研究设计与样本

  • 研究扫描了多能CT体模、拟人头颅体模与尸头,使用研究型PCD-CT以超高清模式、100 kV管电压附加锡滤过(Sn-100 kV)、容积CT剂量指数10 mGy进行,并在商用EID-CT上按标准临床协议扫描。
  • 30例临床指征鼻窦检查患者与2例颞骨检查患者在先完成EID-CT临床扫描后接受PCD-CT扫描。
  • 鼻窦队列PCD-CT采用Sn-100 kV四个剂量水平:10 mGy(n=9)、8 mGy(n=7)、7 mGy(n=7)、6 mGy(n=7);临床EID-CT为120 kV、13.7 mGy(平均CT容积剂量指数)。
  • 颞骨扫描PCD-CT采用Sn-100 kV(10.1 mGy),EID-CT为120 kV、常规临床剂量(52.6与66 mGy)。
  • 鼻窦图像均以H70卷积核、0.75 mm层厚重建,颞骨图像以U70卷积核、0.6 mm层厚重建。

关键影像表现/征象

  • 多能体模实验显示Sn-100 kV PCD-CT图像噪声降低26%,对应相比EID-CT(临床剂量)总剂量降低56%,且未影响图像对比度。
  • 头颅体模与尸头扫描显示PCD-CT相比EID-CT在鼻窦与颞骨检查中分别降低剂量67%与83%。
  • 鼻窦患者队列中PCD-CT平均剂量降低67%。
  • 10与8 mGy的PCD-CT鼻窦患者图像对关键鼻窦结构显示显著优于临床EID-CT:腭小孔中位评分5±0.82、P=0.01;鼻上颌缝中位评分4±0.68、P=0.039;筛前动脉管中位评分4±0.96、P=0.01。
  • 6与7 mGy的PCD-CT鼻窦患者图像与EID-CT之间未显示任何显著差异。

诊断标准与关键数值

  • 鼻窦图像采用H70卷积核、0.75 mm层厚重建,颞骨图像采用U70卷积核、0.6 mm层厚重建。
  • 关键解剖结构显示采用5分Likert量表评估:结构检测不清且诊断信心差=1分;解剖 delineation与诊断信心改善=5分。
  • 空间分辨率改善以线体模测量的调制传递函数量化,V80锐利卷积核的10%调制传递函数=18.6 cm。

临床应用与阅片要点

  • Sn-100 kV PCD-CT在鼻窦检查中10与8 mGy剂量水平对关键结构的显示优于临床EID-CT,6与7 mGy时与EID-CT无显著差异。
  • 颞骨检查中PCD-CT最高降低85%剂量,内耳结构如砧锤关节显示与EID-CT相似。
  • V80锐利卷积核迭代重建可提高鼻上颌缝等结构的可见性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The aim of this study was to quantitatively demonstrate radiation dose reduction for sinus and temporal bone examinations using high-resolution photon-counting detector (PCD) computed tomography (CT) with an additional tin (Sn) filter.

MATERIALS AND METHODS: A multienergy CT phantom, an anthropomorphic head phantom, and a cadaver head were scanned on a research PCD-CT scanner using ultra-high-resolution mode at 100-kV tube potential with an additional tin filter (Sn-100 kV) and volume CT dose index of 10 mGy. They were also scanned on a commercial CT scanner with an energy-integrating detector (EID) following standard clinical protocols. Thirty patients referred to clinically indicated sinus examinations, and two patients referred to temporal bone examinations were scanned on the PCD-CT system after their clinical scans on an EID-CT. For the sinus cohort, PCD-CT scans were performed using Sn-100 kV at 4 dose levels at 10 mGy (n = 9), 8 mGy (n = 7), 7 mGy (n = 7), and 6 mGy (n = 7), and the clinical EID-CT was performed at 120 kV and 13.7 mGy (mean CT volume dose index). For the temporal bone scans, PCD-CT was performed using Sn-100 kV (10.1 mGy), and EID-CT was performed at 120 kV and routine clinical dose (52.6 and 66 mGy). For both PCD-CT and EID-CT, sinus images were reconstructed using H70 kernel at 0.75-mm slice thickness, and temporal bone images were reconstructed using a U70 kernel at 0.6-mm slice thickness. In addition, iterative reconstruction with a dedicated sharp kernel (V80) was used to obtain high-resolution PCD-CT images from a sinus patient scan to demonstrate improved anatomic delineation. Improvements in spatial resolution from the dedicated sharp kernel was quantified using modulation transfer function measured with a wire phantom. A neuroradiologist assessed the H70 sinus images for visualization of critical anatomical structures in low-dose PCD-CT images and routine-dose EID-CT images using a 5-point Likert scale (structural detection obscured and poor diagnostic confidence, score = 1; improved anatomic delineation and diagnostic confidence, score = 5). Image contrast and noise were measured in representative regions of interest and compared between PCD-CT and EID-CT, and the noise difference between the 2 acquisitions was used to estimate the dose reduction in the sinus and temporal bone patient cohorts.

RESULTS: The multienergy phantom experiment showed a noise reduction of 26% in the Sn-100 kV PCD-CT image, corresponding to a total dose reduction of 56% compared with EID-CT (clinical dose) without compromising image contrast. The PCD-CT images from the head phantom and the cadaver scans demonstrated a dose reduction of 67% and 83%, for sinus and temporal bone examinations, respectively, compared with EID-CT. In the sinus cohort, PCD-CT demonstrated a mean dose reduction of 67%. The 10- and 8-mGy sinus patient images from PCD-CT were significantly superior to clinical EID-CT for visualization of critical sinus structures (median score = 5 ± 0.82 and P = 0.01 for lesser palatine foramina, median score = 4 ± 0.68 and P = 0.039 for nasomaxillary sutures, and median score = 4 ± 0.96 and P = 0.01 for anterior ethmoid artery canal). The 6- and 7-mGy sinus patient images did not show any significant difference between PCD-CT and EID-CT. In addition, V80 (sharp kernel, 10% modulation transfer function = 18.6 cm) PCD-CT images from a sinus patient scan increased the conspicuity of nasomaxillary sutures compared with the clinical EID-CT images. The temporal bone patient images demonstrated a dose reduction of up to 85% compared with clinical EID-CT images, whereas visualization of inner ear structures such as the incudomalleolar joint were similar between EID-CT and PCD-CT.

CONCLUSIONS: Phantom and cadaver studies demonstrated dose reduction using Sn-100 kV PCD-CT compared with current clinical EID-CT while maintaining the desired image contrast. Dose reduction was further validated in sinus and temporal bone patient studies. The ultra-high resolution capability from PCD-CT allowed improved anatomical delineation for sinus imaging compared with current clinical standard.