急性缺血性卒中自动ASPECTS:与CT灌注的对比分析
本研究纳入58例前循环卒中并成功行机械取栓的患者,比较自动ASPECTS软件与2名神经放射科医生共识判读及CTP-CBV ASPECTS的一致性。自动ASPECTS与共识判读及CTP-CBV ASPECTS均呈优一致(κ=0.84),且与最终梗死体积负相关(r=-0.66)。作者认为自动ASPECTS可标准化报告并减少判读变异。
要点
- 自动ASPECTS与专家共识判读及CTP-CBV ASPECTS一致性优(κ均=0.84)。
- 自动ASPECTS与最终梗死体积负相关(r=-0.66),优于共识判读(r=-0.52)。
- 按发病时间(>6或≤6小时)分组,自动评分与共识及CTP-CBV相当。
核心结论
- 自动软件得出的ASPECTS与专家神经放射科医生共识判读及同期CTP-CBV ASPECTS表现相当,可用于标准化ASPECTS报告并减少判读变异。
研究设计与样本
- 回顾性纳入58例前循环卒中、有基线NCCT和CTP并成功行机械取栓的患者;NCCT-ASPECTS由2名神经放射科医生评估、分歧共识解决,CTP-CBV ASPECTS由另一名神经放射科医生评估,自动ASPECTS由Brainomix软件提供。
诊断标准与关键数值
- ASPECTS二分为≥6或<6分,并依据发病时间(>6或≤6小时)分组;NCCT ASPECTS观察者间一致性中等(κ=0.48),二分数据略提高(κ=0.64);自动ASPECTS与共识判读(κ=0.84)及CTP-CBV ASPECTS(κ=0.84)一致性优;三组ASPECTS的组内相关系数原始评分0.84(95%CI 0.76-0.90)、二分评分0.94(95%CI 0.91-0.96)。
临床应用与阅片要点
- 自动ASPECTS与专家共识判读及CTP-CBV ASPECTS表现相当,可用于标准化ASPECTS报告并减少判读变异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Automated ASPECTS has the potential of reducing interobserver variability in the determination of early ischemic changes. We aimed to assess the performance of an automated ASPECTS software against the assessment of a neuroradiologist in a comparative analysis with concurrent CTP-based CBV ASPECTS.
MATERIALS AND METHODS: Patients with anterior circulation stroke who had baseline NCCT and CTP and underwent successful mechanical thrombectomy were included. NCCT-ASPECTS was assessed by 2 neuroradiologists, and discrepancies were resolved by consensus. CTP-CBV ASPECTS was assessed by a different neuroradiologist. Automated ASPECTS was provided by Brainomix software. ASPECTS was dichotomized (ASPECTS ≥6 or <6) and was also based on the time from onset (>6 or ≤6 hours).
RESULTS: A total of 58 patients were included. The interobserver agreement for NCCT ASPECTS was moderate (κ = 0.48) and marginally improved (κ = 0.64) for dichotomized data. Automated ASPECTS showed excellent agreement with consensus reads (κ = 0.84) and CTP-CBV ASPECTS (κ = 0.84). Intraclass correlation coefficients for ASPECTS across all 3 groups were 0.84 (95% CI, 0.76-0.90, raw scores) and 0.94 (95% CI, 0.91-0.96, dichotomized scores). Automated scores were comparable with consensus reads and CTP-CBV ASPECTS in patients when grouped on the basis of time from symptom onset (>6 or ≤6 hours). There was significant (P < .001) negative correlation with final infarction volume and the 3 ASPECTS groups (r = -0.52, consensus reads; -0.58, CTP-CBV; and -0.66, automated).
CONCLUSIONS: ASPECTS derived from an automated software performs equally as well as consensus reads of expert neuroradiologists and concurrent CTP-CBV ASPECTS and can be used to standardize ASPECTS reporting and minimize interpretation variability.
引用
- Sundaram VK, Goldstein J, Wheelwright D, et al. Automated ASPECTS in Acute Ischemic Stroke: A Comparative Analysis with CT Perfusion. AJNR Am J Neuroradiol. 2019. PMID: 31727750. doi:10.3174/ajnr.A6303