三维双回波滑动反转恢复超短回波时间(DESIRE UTE)MRI全脑髓鞘成像
研究开发了三维双回波滑动反转恢复超短回波时间序列,通过单次扫描生成不同反转时间图像,以第二回波最小信号确定最佳反转时间抑制长T2信号,相减获得髓鞘图像。在体模、尸脑、健康志愿者及多发性硬化患者中验证,多发性硬化病灶归一化信号强度较正常表现白质显著降低。
要点
- DESIRE UTE序列可在临床3-T扫描仪上生成全脑髓鞘图像。
- 多发性硬化病灶归一化信号强度较正常表现白质显著降低(0.19±0.10 vs 0.76±0.11,P<.001)。
- 最佳反转时间通过第二回波最小信号确定,以抑制长T2白质信号。
核心结论
- 双回波滑动反转恢复超短回波时间序列可在临床3-T扫描仪上特异性地生成全脑髓鞘图像。
研究设计与样本
- 前瞻性研究,纳入20名健康志愿者(平均年龄40岁±13,10名女性)和20名多发性硬化患者(平均年龄58岁±8,15名女性),于2017年11月至2019年2月期间接受MRI检查;另在体模和尸脑标本中验证。
关键影像表现/征象
- 提取的髓鞘囊泡具有高信号强度及相应的T2*和T1,为USE序列直接成像超短T2髓鞘质子提供证据。
- 钆贝葡胺体模在宽T1值范围内被DESIRE UTE选择性抑制。
- 离体多发性硬化脑标本中,多发性硬化病灶内观察到信号丢失,并与组织学分析一致。
- 在体研究中,多发性硬化病灶归一化信号强度较正常表现白质显著降低(0.19±0.10 vs 0.76±0.11,P<.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Signal contamination from long T2 water is a major challenge in direct imaging of myelin with MRI. Nulling of the unwanted long T2 signals can be achieved with an inversion recovery (IR) preparation pulse to null long T2 white matter within the brain. The remaining ultrashort T2 signal from myelin can be detected with an ultrashort echo time (UTE) sequence. Purpose To develop patient-specific whole-brain myelin imaging with a three-dimensional double-echo sliding inversion recovery (DESIRE) UTE sequence. Materials and Methods The DESIRE UTE sequence generates a series of IR images with different inversion times during a single scan. The optimal inversion time for nulling long T2 signal is determined by finding minimal signal on the second echo. Myelin images are generated by subtracting the second echo image from the first UTE image. To validate this method, a prospective study was performed in phantoms, cadaveric brain specimens, healthy volunteers, and patients with multiple sclerosis (MS). A total of 20 healthy volunteers (mean age, 40 years ± 13 [standard deviation], 10 women) and 20 patients with MS (mean age, 58 years ± 8; 15 women) who underwent MRI between November 2017 and February 2019 were prospectively included. Analysis of variance was performed to evaluate the signal difference between MS lesions and normal-appearing white matter in patients with MS. Results High signal intensity and corresponding T2* and T1 of the extracted myelin vesicles provided evidence for direct imaging of ultrashort-T2 myelin protons using the UTE sequence. Gadobenate dimeglumine phantoms with a wide range of T1 values were selectively suppressed with DESIRE UTE. In the ex vivo brain study of MS lesions, signal loss was observed in MS lesions and was conformed with histologic analysis. In the human study, there was a significant reduction in normalized signal intensity in MS lesions compared with that in normal-appearing white matter (0.19 ± 0.10 vs 0.76 ± 0.11, respectively; P < .001). Conclusion The double-echo sliding inversion recovery ultrashort echo time sequence can generate whole-brain myelin images specifically with a clinical 3-T scanner. © RSNA, 2019 Online supplemental material is available for this article. See also the editorial by Port in this issue.
引用
- Ma YJ, Searleman AC, Jang H, et al. Whole-Brain Myelin Imaging Using 3D Double-Echo Sliding Inversion Recovery Ultrashort Echo Time (DESIRE UTE) MRI. Radiology. 2019. PMID: 31746689. doi:10.1148/radiol.2019190911
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。