文献速递泌尿生殖骨肌技术与安全CT

骨科金属伪影减少算法有助于髋关节假体术后泌尿系CT评估

Orthopaedic metallic artefact reduction algorithm facilitates CT evaluation of the urinary tract after hip prosthesis.
PubMed 文献 · Clin Radiol · 2019-10-04

回顾性纳入34例因非假体相关原因行盆腔CT的患者,比较标准迭代重建(ITR)与骨科金属伪影减少算法(O-MAR)对假体侧输尿管和膀胱病变显示的影响。O-MAR显著降低假体侧输尿管和膀胱密度,改善输尿管可见度,并在4例(13%)中改变了诊断。

要点

  • 假体侧输尿管在ITR上53.7%被遮蔽,O-MAR仅4.9%被遮蔽。
  • O-MAR使假体侧输尿管平均密度从-94.76 HU升至-13.40 HU,差异有统计学意义(p<0.0004)。
  • 4例(13%)O-MAR图像显著改变诊断,发现输尿管结石2例、膀胱结石1例、膀胱肿瘤1例。

核心结论

  • O-MAR重建可显著改善金属髋假体邻近泌尿道的可见度,作者认为O-MAR对于检测髋假体患者的输尿管和膀胱病变是必要的。

研究设计与样本

  • 回顾性纳入34例因非假体相关原因行盆腔CT的患者,同时重建标准迭代重建(ITR)和骨科金属伪影减少算法(O-MAR)图像,比较假体侧输尿管和膀胱的密度及半定量可见度评分。

关键影像表现/征象

  • 假体侧输尿管平均密度在ITR图像为-94.76±150.48 HU,在O-MAR图像为-13.40±36.37 HU(p<0.0004)。
  • 假体侧膀胱壁平均密度在ITR图像为-138.62±182.64 HU,在O-MAR图像为-35.55±40.21 HU(p<0.0003)。
  • 假体侧输尿管可见度:ITR序列53.7%被遮蔽,O-MAR序列4.9%被遮蔽。
  • O-MAR在各评估层面的可见度评分均更优(p<0.001)。

临床应用与阅片要点

  • 对髋假体患者行盆腔CT评估输尿管和膀胱时,O-MAR重建可减少金属伪影,改善假体邻近结构的可见度。
  • O-MAR图像有助于发现被伪影遮蔽的输尿管结石、膀胱结石和膀胱肿瘤(本组4例,13%)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To examine the improvement in the visualisation of bladder and ureteric pathologies next to a hip prosthesis with metallic artefact reduction for orthopaedic implants (O-MAR).

MATERIALS AND METHODS: Thirty-four patients who underwent pelvic computed tomography (CT) for non-prosthesis-related causes were enrolled retrospectively. Portal venous phase scans were reconstructed both with standard iterative reconstruction (ITR) and with O-MAR. The density of the ureters and the bladder was measured at both sides in the plane of the prosthesis. A semi-quantitative score was also used to assess visibility. The R (version 3.4.1) package was used for statistical analysis.

RESULTS: The average (μ) density of the 41 prosthesis side ureters was significantly lower on ITR images (μ=-94.76±150.48 [±SD] HU) than on O-MAR images (μ=-13.40±36.37 HU; p<0.0004). The difference between the ITR and O-MAR (μ=-138.62±182.64 versus -35.55±40.21 HU; p<0.0003) was also significant at the prosthesis side of the bladder. The visibility of the prosthesis side ureters was improved: 53.7% was obscured on ITR series compared to 4.9% on O-MAR. The visibility score was also better across all levels (p<0.001) with O-MAR. In four cases (13%), the O-MAR images significantly changed the diagnosis: in two cases ureteric stones, in one case each a bladder stone and a bladder tumour were discovered.

CONCLUSIONS: O-MAR reconstruction of CT images significantly improves the visibility of the urinary tract adjacent to metallic hip implants. Thus, O-MAR is essential for detecting ureteric and bladder pathologies in patients with a hip prosthesis.

来源:Clin Radiol,查看原文。版权归原作者所有。

引用

  1. Kaposi PN, Youn T, Tóth A, et al. Orthopaedic metallic artefact reduction algorithm facilitates CT evaluation of the urinary tract after hip prosthesis. Clin Radiol. 2019. PMID: 31590913. doi:10.1016/j.crad.2019.08.027

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。