不同体型患者使用光谱探测器CT行腹部CT检查的辐射剂量与图像质量:体模与临床研究
本研究在体模和临床患者中比较光谱探测器CT(SDCT)与常规单能CT行腹部CT的辐射剂量与图像质量。体模中SDCT的CTDIvol和噪声均高于常规CT,相对差异分别为7.8%(范围5.3-14%)和平均9.0%(范围5.5-11%)。患者研究中,SDCT在较小患者中CTDIvol趋于更低(200 mm水当量直径时-2%,95%CI -5%至-0.2%),在较大患者中更高(400 mm时8%,95%CI 6%至11%);除门静脉噪声更高和主动脉信号更高外,两机图像质量指标无差异。
要点
- SDCT在200 mm水当量直径患者中CTDIvol较常规CT低2%(95%CI -5%至-0.2%),在400 mm患者中高8%(95%CI 6%至11%)。
- 体模中SDCT的CTDIvol相对常规CT高7.8%(范围5.3-14%),噪声平均高9.0%(范围5.5-11%)。
- 除SDCT上门静脉噪声更高、主动脉信号更高外,两机主观图像质量、SNR、CNR及图像噪声无差异。
核心结论
- SDCT行腹部CT的辐射剂量在平均体型患者中与常规CT相似,在较小患者中更低,在较大患者中略高;两机图像质量相似。
研究设计与样本
- 体模研究使用4个半拟人体模在SDCT(IQon, Philips Healthcare)和匹配扫描参数的常规单能CT(iCT 256, Philips Healthcare)上扫描;临床部分为经IRB批准、符合HIPAA的回顾性队列研究,纳入718例SDCT患者和937例常规CT患者,按患者水当量直径调整后比较辐射剂量,并在50例先后使用两机扫描的患者子集中由两名读者评估主观和客观图像质量。
关键影像表现/征象
- SDCT上门静脉图像噪声高于常规CT,主动脉信号高于常规CT。
- 两机的主观图像质量、SNR、CNR和图像噪声无差异。
诊断标准与关键数值
- 体模中SDCT的CTDIvol相对常规CT高7.8%(范围5.3-14%),噪声平均高9.0%(范围5.5-11%)。
- 患者研究中,200 mm水当量直径时SDCT的CTDIvol较常规CT低2%(95%CI -5%至-0.2%),400 mm时高8%(95%CI 6%至11%)。
临床应用与阅片要点
- 对于较小体型患者,SDCT腹部CT的辐射剂量可能低于常规CT,而对于较大体型患者则略高。
- 阅片时注意SDCT上门静脉噪声更高、主动脉信号更高,但整体图像质量与常规CT相似。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare radiation dose and image quality for abdominal CTs performed on a spectral detector CT (SDCT) and a comparable single-energy conventional CT scanner for patients of different sizes.
METHODS: Four semi-anthropomorphic phantoms were scanned on an SDCT (IQon, Philips Healthcare) and a comparable single-energy CT (iCT 256, Philips Healthcare) under matched scan parameters. Image noise and radiation dose were compared. For the HIPAA-compliant, IRB-approved retrospective cohort patient study, radiation dose was compared after adjusting for patient water equivalent diameter. Difference in subjective and objective image quality was assessed on a subset of 50 patients scanned on both scanners by two readers.
RESULTS: CTDIvol and noise from SDCT were higher than conventional CT for all phantoms, with a relative difference of 7.8% (range 5.3-14%) for radiation dose and average difference of 9.0% (range 5.5-11%) for noise. 718 SDCT and 937 conventional CT patients were included in the patient study. CTDIvol for SDCT patients tends to be lower for smaller patients (- 2%, 95% confidence interval (- 5%, - 0.2%) for 200 mm water equivalent diameter) and higher for larger patients compared to conventional CT (8%, (6%, 11%) for 400 mm). No difference was seen for subjective image quality, SNR, CNR, or image noise between the two scanners, except for higher image noise in the portal vein and higher signal in the aorta on SDCT.
CONCLUSION: Radiation dose for abdominal CT performed on SDCT is similar to the dose on a conventional CT for average size patients, lower for smaller patients, and slightly higher for larger patients. Image quality is similar between the two scanners.
引用
- Duan X, Ananthakrishnan L, Guild JB, et al. Radiation doses and image quality of abdominal CT scans at different patient sizes using spectral detector CT scanner: a phantom and clinical study. Abdom Radiol (NY). 2020. PMID: 31587100. doi:10.1007/s00261-019-02247-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。