腹部CT中分滤双束双能量与单能量图像质量的比较
前瞻性研究纳入103例患者,分滤双束双能量(TBDE,n=51)或单能量CT(SECT,n=52)扫描,比较腹部六种结构的客观与主观图像质量。TBDE所有结构SNR更高(p<0.05),CT值两组可比(p>0.05),噪声降低7-17%(p<0.01)且降幅随体型增大,主观某些方面TBDE更优。
要点
- TBDE在相同辐射水平下可获得与SECT相似或更优的客观及主观图像质量。
- TBDE图像噪声较SECT降低7-17%(p<0.01),且噪声降低幅度随扫描部位体型增大而增加。
- 两组间各结构CT值可比(p>0.05),TBDE所有结构SNR均高于SECT(p<0.05)。
核心结论
- TBDE扫描模式在相同辐射水平下可产生与常规SECT相似甚至更好的客观和主观图像质量。
- 定量上,TBDE图像噪声降低7-17%,具体取决于所扫描身体区域的大小。
研究设计与样本
- 前瞻性研究,共103例患者,TBDE组n=51、SECT组n=52。
- 测量肝、脾、脂肪、肌肉、主动脉和门静脉六个腹部结构的CT值和噪声,并单独测量肝脏归一化噪声水平。
- 比较两组间SNR和CT值的一致性;由两名放射科医生以盲法从六个方面评估主观图像质量,采用Cohen Kappa统计量评价观察者间一致性。
诊断标准与关键数值
- 客观图像质量:TBDE所有结构SNR均高于SECT(p<0.05)。
- 不同结构CT值两组间可比(p>0.05)。
- 所有患者体型中,TBDE图像噪声较SECT显著降低(降低7-17%,p<0.01),且噪声降低幅度随体型增大而增加。
- 主观图像质量:TBDE图像在某些方面更优;两名观察者间Cohen Kappa值为0.7634-0.8460,提示一致性充分。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare the objective and subjective image quality between composed images from split-filter twin beam dual energy (TBDE) and single-energy computed tomography (SECT) in abdominal CT.
METHODS: In this prospective study, 103 patients were imaged using TBDE (n = 51) or SECT (n = 52). The CT number and noise were measured for the following six abdominal structures: liver, spleen, fat, muscle, aorta and portal vein. The normalised noise level for the liver was separately measured and compared. The consistency of the SNR and CT number was compared between the two groups. The subjective image quality was evaluated using six aspects in a blinded manner. Cohen's Kappa statistic was used to determine the level of agreement between the two radiologists.
RESULTS: For the objective image quality comparison, the SNR of all structures was higher using TBDE compared to SECT (p < 0.05). The CT value for different structures were comparable between the two groups (p > 0.05). Among all patient sizes, the noise level for TBDE images was significantly lower (7-17% reduction) compared to the SECT images (p < 0.01). Furthermore, noise reduction's magnitude increases with body size. For image quality's subjective evaluation, TBDE images are superior for certain aspects. Cohen's Kappa values (0.7634-0.8460) suggest an adequate level of agreement between the two observers.
CONCLUSIONS: TBDE scan mode can yield similar or even better objective and subjective image quality at the same level of radiation than conventional SECT. Quantitatively, TBDE images have a 7-17% reduction in noise, depending on the size of the scanned body regions.
引用
- Niu Z, Chen J, Ren H, et al. Comparison of image quality between split-filter twin beam dual energy and single energy images in abdominal CT. Eur J Radiol. 2019. PMID: 31648102. doi:10.1016/j.ejrad.2019.108702
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。