卷积神经网络运动校正冠状动脉钙化评分:机器人模拟研究
本研究用含9个不同密度和大小钙化斑块的3条人工冠状动脉附着于运动机械臂,在0–90 mm/s运动下扫描生成运动伪影图像,采用Inception v3卷积神经网络分类并校正Agatston评分。CNN分类总体准确率79.2±6.1%,校正后Agatston评分变异从37.8%降至3.7%(p<0.001),钙化检出敏感度从65%提高至85%。
要点
- CNN可对运动模糊图像分类并校正非触发胸部CT的钙化评分,总体分类准确率79.2±6.1%。
- CNN校正使运动斑块Agatston评分变异从37.8%降至3.7%(p<0.001)。
- CNN校正使钙化检出敏感度(非零评分比例)从65%提高至85%。
核心结论
- 实验研究表明,CNN能够分类运动诱导的模糊图像并校正非触发胸部CT获得的钙化评分。
- CNN校正大幅降低总体Agatston评分变异并提高钙化检出敏感度。
研究设计与样本
- 实验研究:3条人工冠状动脉含9个不同密度(高、中、低)和大小(大、中、小)的钙化斑块,附着于运动机械臂。
- 人工动脉以0–90 mm/s运动扫描,生成9类(每个钙化斑块一类)含运动伪影的图像。
- 对Inception v3 CNN进行微调与验证,以CNN预测分类对应的Agatston评分作为校正评分。
- 以静止状态评分为参考,计算运动斑块Agatston评分变异及CNN校正后的变异。
诊断标准与关键数值
- CNN分类9类的总体准确率为79.2±6.1%。
- 高密度斑块分类准确率88.3±4.9%,中密度75.9±6.4%,低密度73.5±5.0%。
- 与静止Agatston评分相比,运动斑块总体中位评分变异为37.8%(第1和第3四分位数10.5%和68.8%)。
- CNN校正后变异降至3.7%(1.9%,9.1%)(p<0.001,Mann-Whitney U检验)。
- 钙化检出敏感度(全部评分中非零评分的百分比)从65%提高至85%。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To classify motion-induced blurred images of calcified coronary plaques so as to correct coronary calcium scores on nontriggered chest CT, using a deep convolutional neural network (CNN) trained by images of motion artifacts.
METHODS: Three artificial coronary arteries containing nine calcified plaques of different densities (high, medium, and low) and sizes (large, medium, and small) were attached to a moving robotic arm. The artificial arteries moving at 0-90 mm/s were scanned to generate nine categories (each from one calcified plaque) of images with motion artifacts. An inception v3 CNN was fine-tuned and validated. Agatston scores of the predicted classification by CNN were considered as corrected scores. Variation of Agatston scores on moving plaque and by CNN correction was calculated using the scores at rest as reference.
RESULTS: The overall accuracy of CNN classification was 79.2 ± 6.1% for nine categories. The accuracy was 88.3 ± 4.9%, 75.9 ± 6.4%, and 73.5 ± 5.0% for the high-, medium-, and low-density plaques, respectively. Compared with the Agatston score at rest, the overall median score variation was 37.8% (1st and 3rd quartile, 10.5% and 68.8%) in moving plaques. CNN correction largely decreased the variation to 3.7% (1.9%, 9.1%) (p < 0.001, Mann-Whitney U test) and improved the sensitivity (percentage of non-zero scores among all the scores) from 65 to 85% for detection of coronary calcifications.
CONCLUSIONS: In this experimental study, CNN showed the ability to classify motion-induced blurred images and correct calcium scores derived from nontriggered chest CT. CNN correction largely reduces the overall Agatston score variation and increases the sensitivity to detect calcifications.
KEY POINTS: • A deep CNN architecture trained by CT images of motion artifacts showed the ability to correct coronary calcium scores from blurred images. • A correction algorithm based on deep CNN can be used for a tenfold reduction in Agatston score variations from 38 to 3.7% of moving coronary calcified plaques and to improve the sensitivity from 65 to 85% for the detection of calcifications. • This experimental study provides a method to improve its accuracy for coronary calcium scores that is a fundamental step towards a real clinical scenario.
引用
- Zhang Y, van der Werf NR, Jiang B, et al. Motion-corrected coronary calcium scores by a convolutional neural network: a robotic simulating study. Eur Radiol. 2019. PMID: 31630233. doi:10.1007/s00330-019-06447-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。