儿童头部CT采用先进模型迭代重建:聚焦图像质量与辐射剂量降低
本研究评估儿童头部CT采用先进模型迭代重建(ADMIRE)后的图像质量。将剂量降低后的ADMIRE重建图像与常规滤波反投影(cFBP)组比较,结果显示管电流降低后CT剂量指数体积平均降低30.6%、剂量长度乘积降低32.1%、有效剂量降低32.1%,且ADMIRE强度级别1至5级可逐步降低噪声并提高信噪比和对比噪声比,结论为ADMIRE可在不降低图像质量的前提下降低辐射剂量。
要点
- 儿童头部CT采用ADMIRE可在降低约30%辐射剂量的同时保持图像质量。
- ADMIRE强度级别从1到5递增时,噪声逐步降低,信噪比和对比噪声比逐步升高。
- ADMIRE各级别的主观图像质量参数也随级别递增而逐步改善。
核心结论
- 儿童头部CT使用ADMIRE可降低辐射剂量而不降低图像质量。
研究设计与样本
- 研究纳入常规滤波反投影重建组(cFBP组,n=105)和剂量降低后同时采用滤波反投影与ADMIRE重建的低剂量组(n=109),低剂量组适配全部五个ADMIRE强度级别并与cFBP组比较。
诊断标准与关键数值
- 管电流降低后,CT剂量指数体积平均降低30.6%,剂量长度乘积平均降低32.1%,有效剂量平均降低32.1%。
- ADMIRE强度级别从1到5递增时,空气、脑脊液和白质中的噪声逐步降低(P<0.001)。
- 所有年龄组的信噪比和对比噪声比随ADMIRE强度级别从1到5递增而升高,差异有统计学意义(P<0.001)。
- 定性参数(噪声、幕上和幕下白质与灰质分化、锐利度、伪影及诊断准确性)随ADMIRE强度级别从1到5递增而逐步改善(P<0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Iterative reconstruction has become the standard method for reconstructing computed tomography (CT) scans and needs to be verified for adaptation.
OBJECTIVE: To assess the image quality after adapting advanced modeled iterative reconstruction (ADMIRE) for pediatric head CT.
MATERIALS AND METHODS: We included image sets with filtered back projection reconstruction (the cFBP group, n=105) and both filtered back projection and ADMIRE reconstruction (the lower-dose group, n=109) after dose reduction. All five strength levels of ADMIRE and filtered back projection were adapted for the lower-dose group and compared with the cFBP group. Quantitative parameters including noise, signal-to-noise ratio and contrast-to-noise ratio and qualitative parameters including noise, white matter and gray matter differentiation of the supra- and infratentorial levels, sharpness, artifact, and diagnostic accuracy were also evaluated and compared with interobserver agreement.
RESULTS: There was a mean dose reduction of 30.6% in CT dose index volume, 32.1% in dose length product, and 32.1% in effective dose after tube current reduction. There was gradual reduction of noise in air, cerebrospinal fluid and white matter with strength levels of ADMIRE from 1 to 5 (P<0.001). Signal-to-noise ratio and contrast-to-noise ratio in all age groups increased among strength levels of ADMIRE, in sequence from 1 to 5, with statistical significance (P<0.001). Gradual reduction of qualitative parameters was noted among strength levels of ADMIRE in sequence from 1 to 5 (P<0.001).
CONCLUSION: Use of ADMIRE for pediatric head CT can reduce radiation dose without degrading image quality.
引用
- Cho HH, Lee SM, You SK, et al. Pediatric head computed tomography with advanced modeled iterative reconstruction: focus on image quality and reduction of radiation dose. Pediatr Radiol. 2019. PMID: 31630218. doi:10.1007/s00247-019-04532-z
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。