文献速递神经技术与安全CT

低剂量螺旋头部CT检测导致神经功能缺损的颅内病变的评价

Evaluation of Lower-Dose Spiral Head CT for Detection of Intracranial Findings Causing Neurologic Deficits.
PubMed 文献 · AJNR Am J Neuroradiol · 2019-10-24

研究纳入83例疑似神经功能缺损患者,比较常规剂量(250有效mAs)与低剂量(25-200有效mAs)螺旋头部CT对颅内病变的检出效能。以jackknife替代自由响应受试者工作特征图优度为指标,100有效mAs迭代重建和200有效mAs滤波反投影非劣效于常规剂量,迭代重建在平均CT剂量指数15.2 mGy时保持非劣效。

要点

  • 100有效mAs迭代重建和200有效mAs滤波反投影的检出效能非劣效于250有效mAs迭代重建。
  • 100有效mAs滤波反投影及更低剂量水平未达到非劣效。
  • 图像质量随剂量升高和迭代重建而更优(P < .05)。

核心结论

  • 100-200有效mAs剂量水平的观察者效能非劣效于250有效mAs迭代重建,迭代重建在平均CT剂量指数15.2 mGy时保持非劣效。

研究设计与样本

  • 收集83例因疑似神经功能缺损行平扫螺旋头部CT患者的投影数据;常规剂量采用250有效mAs迭代重建,低剂量配置包括25有效mAs迭代重建、50有效mAs滤波反投影和迭代重建、100有效mAs滤波反投影和迭代重建、200有效mAs滤波反投影;由3名神经放射学家勾画病变并评估诊断、置信度(0-100)和图像质量。

诊断标准与关键数值

  • 阳性病例(梗死、轴内出血、肿块或轴外出血)需经组织病理学、手术、病变进展或相应神经功能缺损确认;阴性病例需经两名神经放射学家和一名临床神经学家评估为阴性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Despite the frequent use of unenhanced head CT for the detection of acute neurologic deficit, the radiation dose for this exam varies widely. Our aim was to evaluate the performance of lower-dose head CT for detection of intracranial findings resulting in acute neurologic deficit.

MATERIALS AND METHODS: Projection data from 83 patients undergoing unenhanced spiral head CT for suspected neurologic deficits were collected. Cases positive for infarction, intra-axial hemorrhage, mass, or extra-axial hemorrhage required confirmation by histopathology, surgery, progression of findings, or corresponding neurologic deficit; cases negative for these target diagnoses required negative assessments by two neuroradiologists and a clinical neurologist. A routine dose head CT was obtained using 250 effective mAs and iterative reconstruction. Lower-dose configurations were reconstructed (25-effective mAs iterative reconstruction, 50-effective mAs filtered back-projection and iterative reconstruction, 100-effective mAs filtered back-projection and iterative reconstruction, 200-effective mAs filtered back-projection). Three neuroradiologists circled findings, indicating diagnosis, confidence (0-100), and image quality. The difference between the jackknife alternative free-response receiver operating characteristic figure of merit at routine and lower-dose configurations was estimated. A lower 95% CI estimate of the difference greater than -0.10 indicated noninferiority.

RESULTS: Forty-two of 83 patients had 70 intracranial findings (29 infarcts, 25 masses, 10 extra- and 6 intra-axial hemorrhages) at routine head CT (CT dose index = 38.3 mGy). The routine-dose jackknife alternative free-response receiver operating characteristic figure of merit was 0.87 (95% CI, 0.81-0.93). Noninferiority was shown for 100-effective mAs iterative reconstruction (figure of merit difference, -0.04; 95% CI, -0.08 to 0.004) and 200-effective mAs filtered back-projection (-0.02; 95% CI, -0.06 to 0.02) but not for 100-effective mAs filtered back-projection (-0.06; 95% CI, -0.10 to -0.02) or lower-dose levels. Image quality was better at higher-dose levels and with iterative reconstruction (P < .05).

CONCLUSIONS: Observer performance for dose levels using 100-200 eff mAs was noninferior to that observed at 250 effective mAs with iterative reconstruction, with iterative reconstruction preserving noninferiority at a mean CT dose index of 15.2 mGy.