高性能低场强MRI在介入与诊断成像中的机遇
研究将商用1.5-T MRI系统改造为0.55 T并保持高性能硬件,评估其在金属器械心血管导管介入、高磁化率区域诊断成像及高效图像采集中的应用。在83例0.55-T检查中,T1缩短32%、T2延长26%、T2*延长40%,9种金属介入器械在0.55 T下产热<1°C,7例受试者完成MRI引导右心导管术;氧吸入使肺信号增强19%±11 vs 7.6%±6.3(1.5 T,P=.02)。
要点
- 0.55 T下金属器械产热<1°C,9种商用金属器械可安全用于MRI引导右心导管术。
- 低场强改善磁场均匀性,肺、上气道、颅窦和肠道图像畸变较1.5 T减少。
- 氧吸入肺信号增强在0.55 T为19%±11,高于1.5 T的7.6%±6.3(P=.02)。
核心结论
- 这项初步研究提示,高性能低场强MRI为金属器械MRI引导导管术、高磁化率区域成像和高效成像提供了优势。
研究设计与样本
- 将商用1.5-T MRI系统改造为0.55 T并保持高性能硬件、屏蔽梯度(45 mT/m;200 T/m/sec)和先进成像方法;2018年1月至2019年4月行83例0.55-T MRI检查(45名女性;平均年龄34岁±13)。
关键影像表现/征象
- 与1.5 T相比,0.55 T下T1平均缩短32%、T2平均延长26%、T2*平均延长40%。
- 0.55 T下肺、上气道、颅窦和肠道图像畸变较1.5 T减少,归因于磁场均匀性改善。
- 氧吸入在0.55 T产生肺信号增强19%±11,1.5 T为7.6%±6.3(P=.02;5名受试者),归因于氧的T1弛豫率增加(4.7e-4 mmHg-1sec-1)。
- 螺旋采集适用于低场强,信噪比较笛卡尔采集提高(P<.02)。
- 脑、脊柱、腹部和心脏代表性成像在该系统上获得良好图像质量。
诊断标准与关键数值
- 9种金属介入器械在0.55 T下产热<1°C,判定为本质安全。
- 7名受试者使用商用金属导丝完成MRI引导右心导管术。
临床应用与阅片要点
- 低场强0.55 T配合高性能硬件可用于金属器械MRI引导心血管导管术,器械产热<1°C。
- 在高磁化率区域(肺、上气道、颅窦、肠道)成像时,0.55 T较1.5 T减少图像畸变。
- 螺旋采集在低场强下可获得较笛卡尔采集更高的信噪比(P<.02)。
局限性
- 原文为初步研究(initial study),结论为提示性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Commercial low-field-strength MRI systems are generally not equipped with state-of-the-art MRI hardware, and are not suitable for demanding imaging techniques. An MRI system was developed that combines low field strength (0.55 T) with high-performance imaging technology. Purpose To evaluate applications of a high-performance low-field-strength MRI system, specifically MRI-guided cardiovascular catheterizations with metallic devices, diagnostic imaging in high-susceptibility regions, and efficient image acquisition strategies. Materials and Methods A commercial 1.5-T MRI system was modified to operate at 0.55 T while maintaining high-performance hardware, shielded gradients (45 mT/m; 200 T/m/sec), and advanced imaging methods. MRI was performed between January 2018 and April 2019. T1, T2, and T2* were measured at 0.55 T; relaxivity of exogenous contrast agents was measured; and clinical applications advantageous at low field were evaluated. Results There were 83 0.55-T MRI examinations performed in study participants (45 women; mean age, 34 years ± 13). On average, T1 was 32% shorter, T2 was 26% longer, and T2* was 40% longer at 0.55 T compared with 1.5 T. Nine metallic interventional devices were found to be intrinsically safe at 0.55 T (<1°C heating) and MRI-guided right heart catheterization was performed in seven study participants with commercial metallic guidewires. Compared with 1.5 T, reduced image distortion was shown in lungs, upper airway, cranial sinuses, and intestines because of improved field homogeneity. Oxygen inhalation generated lung signal enhancement of 19% ± 11 (standard deviation) at 0.55 T compared with 7.6% ± 6.3 at 1.5 T (P = .02; five participants) because of the increased T1 relaxivity of oxygen (4.7e-4 mmHg-1sec-1). Efficient spiral image acquisitions were amenable to low field strength and generated increased signal-to-noise ratio compared with Cartesian acquisitions (P < .02). Representative imaging of the brain, spine, abdomen, and heart generated good image quality with this system. Conclusion This initial study suggests that high-performance low-field-strength MRI offers advantages for MRI-guided catheterizations with metal devices, MRI in high-susceptibility regions, and efficient imaging. © RSNA, 2019 Online supplemental material is available for this article. See also the editorial by Grist in this issue.
引用
- Campbell-Washburn AE, Ramasawmy R, Restivo MC, et al. Opportunities in Interventional and Diagnostic Imaging by Using High-Performance Low-Field-Strength MRI. Radiology. 2019. PMID: 31573398. doi:10.1148/radiol.2019190452
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。