文献速递神经技术与安全CT

光谱探测器CT的虚拟单能图像可在头颅平扫CT中实现辐射剂量降低

Virtual Monoenergetic Images from Spectral Detector CT Enable Radiation Dose Reduction in Unenhanced Cranial CT.
PubMed 文献 · AJNR Am J Neuroradiol · 2019-09-19

该回顾性研究纳入140例患者共171次头颅光谱探测器CT检查,在管电流时间乘积从320降至260 mAs(容积CT剂量指数55降至44.7 mGy)条件下,比较65 keV虚拟单能图像与常规图像的噪声、对比噪声比及主观图像质量,发现65 keV虚拟单能图像噪声更低、对比噪声比更高,主观图像质量更优,支持辐射剂量降低。

要点

  • 65 keV虚拟单能图像噪声显著低于常规图像,且不依赖容积CT剂量指数。
  • 65 keV虚拟单能图像在44.7 mGy时的噪声和对比噪声比优于常规图像55 mGy。
  • 主观分析证实65 keV虚拟单能图像在44.7 mGy时图像质量仍更好。

核心结论

  • 光谱探测器CT的65 keV虚拟单能图像可实现头颅CT辐射剂量降低;本概念验证包含19%的辐射剂量降低,数据提示更大降低幅度似乎可实现。

研究设计与样本

  • 回顾性纳入2017年2月至11月间140例连续患者共171次头颅光谱探测器CT检查(年龄56±19岁,男/女比例56%/44%);管电流时间乘积降至320、290、260 mAs分别检查61、55、55例,其他扫描参数保持一致;记录容积CT剂量指数,在常规图像与65 keV虚拟单能图像相同位置放置ROI计算对比噪声比,两名对重建技术盲法的放射科医生以5分Likert量表评估图像质量,统计采用ANOVA和Wilcoxon检验校正多重比较。

诊断标准与关键数值

  • 65 keV虚拟单能图像与常规图像噪声比较:眼外肌在49.8 mGy时,65 keV虚拟单能图像3.7±1.3 HU vs 常规图像5.6±1.6 HU,P < .001。
  • 常规图像噪声随辐射剂量降低而轻度增加:眼外肌在常规图像中分别为5.3±1.4、5.6±1.6、6.1±2.1 HU(对应320、290、260 mAs)。
  • 65 keV虚拟单能图像对比噪声比在所有容积CT剂量指数下均优于常规图像(P < .001)。
  • 65 keV虚拟单能图像在44.7 mGy时噪声显著低于且对比噪声比显著高于常规图像55 mGy(P < .001)。

临床应用与阅片要点

  • 阅片时可利用65 keV虚拟单能图像改善灰白质分化,从而在降低辐射剂量条件下维持图像质量。

局限性

  • 本文为概念验证性研究。
  • 本研究所实现的辐射剂量降低幅度为19%。
  • 数据提示更大降低幅度似乎可实现,但尚需进一步研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Our aim was to evaluate whether improved gray-white matter differentiation in cranial CT by means of 65- keV virtual monoenergetic images enables a radiation dose reduction compared to conventional images.

MATERIALS AND METHODS: One hundred forty consecutive patients undergoing 171 spectral detector CTs of the head between February and November 2017 (56 ± 19 years of age; male/female ratio, 56%/44%) were retrospectively included. The tube current-time product was reduced during the study period, resulting in 61, 55, and 55 patients being examined with 320, 290, and 260 mAs, respectively. All other scanning parameters were kept identical. The volume CT dose index was recorded. ROIs were placed in gray and white matter on conventional images and copied to identical positions in 65- keV virtual monoenergetic images. The contrast-to-noise ratio was calculated. Two radiologists blinded to the reconstruction technique evaluated image quality on a 5-point Likert-scale. Statistical assessment was performed using ANOVA and Wilcoxon test adjusted for multiple comparisons.

RESULTS: The mean volume CT dose index was 55, 49.8, and 44.7 mGy using 320, 290, and 260 mAs, respectively. Irrespective of the volume CT dose index, noise was significantly lower in 65- keV virtual monoenergetic images compared with conventional images (65- keV virtual monoenergetic images/conventional images: extraocular muscle with 49.8 mGy, 3.7 ± 1.3/5.6 ± 1.6 HU, P < .001). Noise slightly increased with a reduced radiation dose (eg, extraocular muscle in conventional images: 5.3 ± 1.4/5.6 ± 1.6/6.1 ± 2.1 HU). Overall, the contrast-to-noise ratio in 65- keV virtual monoenergetic images was superior to that in conventional images irrespective of the volume CT dose index (P < .001). Particularly, 65-keV virtual monoenergetic images with 44.7 mGy showed significantly lower noise and a higher contrast-to-noise ratio than conventional images with 55 mGy (P < .001). Subjective analysis confirmed better image quality in 65- keV virtual monoenergetic images, even using 44.7 mGy.

CONCLUSIONS: The 65-keV virtual monoenergetic images from spectral detector CT allow radiation dose reduction in cranial CT. While this proof of concept included a radiation dose reduction of 19%, our data suggest that even greater reduction appears achievable.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。