PubMed 文献 · Eur Radiol · 2019-09-02

单中心回顾性研究评估电磁导航系统(ENS)辅助CT引导下经皮骨水泥内固定螺钉(FICS)置入的技术成功率与螺钉放置精度。共48例(96%)技术成功,置入76枚螺钉,平均近端偏差8.0±4.5 mm、远端偏差7.5±4.4 mm、角度偏差5.4±2°;头尾成角小于20°的螺钉角度精度更高(4.4° vs 6.4°,p=0.02)。作者认为ENS辅助CT是经皮FICS的有效方法。

要点

  • ENS辅助CT引导下经皮FICS置入螺钉技术成功率达96%(48例)。
  • 头尾成角小于20°的螺钉角度偏差更小(4.4° vs 6.4°,p=0.02)。
  • ENS辅助CT可作为CBCT引导经皮FICS的替代方式。

核心结论

  • ENS辅助CT引导是经皮FICS的有效方法,技术成功率高且置入精度优良。
  • ENS辅助CT可在需要较大头尾成角时实现骨盆环和股骨颈的螺钉置入,轨迹范围广。
  • ENS辅助CT可作为CBCT引导经皮FICS的替代方式。

研究设计与样本

  • 单中心回顾性研究,连续纳入接受ENS辅助CT引导下经皮FICS以预防或姑息骨盆或股骨颈骨折的癌症患者。
  • 主要终点为技术成功,次要终点包括螺钉放置精度(近端偏差p、远端偏差d、角度偏差θ)、辐射剂量、CT采集次数、手术时长和并发症。

关键影像表现/征象

  • 共置入76枚螺钉,平均近端偏差p为8.0±4.5 mm,平均远端偏差d为7.5±4.4 mm,平均角度偏差θ为5.4±2°。
  • 头尾成角小于20°的螺钉角度偏差为4.4°,头尾成角大于等于20°的螺钉为6.4°,前者角度精度更高(p=0.02)。

临床应用与阅片要点

  • ENS辅助CT引导下经皮FICS置入螺钉技术成功率为96%(48例)。
  • ENS辅助CT可在需要较大头尾成角时置入螺钉,头尾成角小于20°时角度精度更优(4.4° vs 6.4°,p=0.02)。
  • ENS辅助CT可作为CBCT引导经皮FICS的替代方式。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate electromagnetic navigation system (ENS) for percutaneous fixation by internal cemented screw (FICS) under CT guidance.

BACKGROUND: FICS is a recently developed modality that consists in inserting screws, under imaging guidance, into bone through a minimal skin incision. FICS recently showed good efficacy for the palliation or prevention of pathologic fractures of the pelvic ring and femoral neck.

MATERIALS AND METHODS: In this single-center retrospective study, we reviewed all consecutive cancer patients treated with percutaneous FICS under ENS-assisted CT guidance for the prevention or palliation of pelvic or femoral neck fractures. The primary endpoint was technical success. Secondary endpoints were screw placement accuracy (defined by proximal deviation p, distal deviation d, and angle deviation θ), radiation dose exposure, number of CT acquisitions, duration of procedures, and complications.

RESULTS: Mean duration of FICS procedures was 111 ± 51 min. Mean post-procedure hospitalization length was 2.1 days. Technical success was achieved in 48 cases (96%) with a total of 76 screws inserted. Mean distance p, mean distance d, and mean angle θ were respectively 8.0 ± 4.5 mm, 7.5 ± 4.4 mm, and 5.4 ± 2°. Angle θ accuracy was higher for screws with a craniocaudal angulation of less than 20° (4.4° vs 6.4°, p = 0.02). The mean number of CT acquisitions during procedures was 6.4 ± 3.0. The mean dose length product was 1524 ± 953 mGy cm and the mean dose area product was 12 ± 8 Gy cm2. Five complications occurred in 4 patients.

CONCLUSION: CT guidance assisted by ENS is an effective approach for percutaneous FICS.

KEY POINTS: • ENS-assisted CT enables screw insertion in the pelvic ring and femoral neck, with a wide range of trajectories, even when a significant craniocaudal angulation is required. • ENS-assisted CT can be used as an alternative to CBCT guidance for percutaneous fixation by internal cemented screw. • ENS-assisted CT provides high technical success rate with excellent placement accuracy.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Moulin B, Tselikas L, De Baere T, et al. CT guidance assisted by electromagnetic navigation system for percutaneous fixation by internal cemented screws (FICS). Eur Radiol. 2019. PMID: 31478088. doi:10.1007/s00330-019-06408-0

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。