双能低keV或单能低kV CT检测内漏:主动脉瘤体模的6名阅片者研究

Dual-Energy Low-keV or Single-Energy Low-kV CT for Endoleak Detection?: A 6-Reader Study in an Aortic Aneurysm Phantom.
PubMed 文献 · Invest Radiol · 2020-01-01

在模拟36个内漏的腹主动脉瘤体模中,比较单能低kV图像与双能低keV虚拟单能图像对内漏检测的图像质量、可见度和敏感度。双能低keV图像主动脉对比噪声比更高,但单能低kV图像在噪声特性、主观图像质量、可见度和内漏敏感度上总体更优。

要点

  • 双能低keV虚拟单能图像主动脉对比噪声比显著高于单能低kV图像(P<0.001)。
  • 单能低kV图像主观图像质量和整体可见度优于双能低keV虚拟单能图像(P<0.05)。
  • 中型体模内漏敏感度单能方案A为60.9%、B为62.2%,双能方案C为54.2%、D为49.3%,B与D差异显著(P<0.05)。

核心结论

  • 低keV虚拟单能图像可提高主动脉对比噪声比,但噪声特性、主观图像质量、可见度和内漏敏感度总体以单能低kV图像更优。

研究设计与样本

  • 体模研究,使用模拟36个内漏(2种密度;直径2、4、6 mm)的腹主动脉瘤体模,分中型和大型患者体型;中型体型扫描单能80 kVp(方案A)和100 kVp(方案B)及双能80/Sn150kVp(方案C),大型体型扫描单能80 kVp(方案A)和100 kVp(方案B)及双能90/Sn150kVp(方案D),从C和D重建40 keV和50 keV虚拟单能图像;辐射剂量中型为3 mGy、大型为6 mGy;由6名放射科医师独立评估。

关键影像表现/征象

  • 双能低keV虚拟单能图像噪声幅度更高、纹理更粗糙。

诊断标准与关键数值

  • 中型体模内漏敏感度:方案A为60.9%、方案B为62.2%、方案C为54.2%、方案D为49.3%,方案B与D差异显著(P<0.05)。
  • 大型体模各方案间内漏敏感度无显著差异(P=0.79),方案B为31.4%、方案C为31.7%最高。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to compare image quality, conspicuity, and endoleak detection between single-energy low-kV images (SEIs) and dual-energy low-keV virtual monoenergetic images (VMIs+) in computed tomography angiography of the aorta after endovascular repair.

MATERIALS AND METHODS: An abdominal aortic aneurysm phantom simulating 36 endoleaks (2 densities; diameters: 2, 4, and 6 mm) in a medium- and large-sized patient was used. Each size was scanned using single-energy at 80 kVp (A) and 100 kVp (B), and dual-energy at 80/Sn150kVp for the medium (C) and 90/Sn150kVp for the large size (D). VMIs+ at 40 keV and 50 keV were reconstructed from protocols C and D. Radiation dose was 3 mGy for the medium and 6 mGy for the large size. Objective image quality and normalized noise power spectrum were determined. Subjective image quality, conspicuity, and sensitivity for endoleaks were independently assessed by 6 radiologists. Sensitivity was compared using Marascuilo procedure and Fisher exact test. Conspicuities were compared using Wilcoxon-matched pairs test, analysis of variance, and Tukey test.

RESULTS: The contrast-to-noise-ratio of the aorta was significantly higher for VMI+ compared with SEI (P < 0.001). Noise power spectrum showed a higher noise magnitude and coarser texture in VMI+. Subjective image quality and overall conspicuity was lower for VMI+ compared with SEI (P < 0.05). Sensitivity for endoleaks was overall higher in the medium phantom for SEI (60.9% for A, 62.2% for B) compared with VMI+ (54.2% for C, 49.3% for D) with significant differences between protocols B and D (P < 0.05). In the large phantom, there was no significant difference in sensitivity among protocols (P = 0.79), with highest rates for protocols B (31.4%) and C (31.7%).

CONCLUSIONS: Our study indicates that low-keV VMI+ results in improved contrast-to-noise-ratio of the aorta, whereas noise properties, subjective image quality, conspicuity, and sensitivity for endoleaks were overall superior for SEI.

来源:Invest Radiol,查看原文。版权归原作者所有。