心脏CT心肌细胞外容积定量:两种迭代重建算法效果比较,以MRI为参考标准

Myocardial extracellular volume quantification in cardiac CT: comparison of the effects of two different iterative reconstruction algorithms with MRI as a reference standard.
PubMed 文献 · Eur Radiol · 2019-08-30

回顾性纳入34例患者,比较混合迭代重建(HIR)与含线束硬化模型的模型迭代重建(MBIR)在心脏CT心肌ECV定量中的效果,以MRI为参考标准。MBIR的CT与MRI ECV平均绝对差显著小于HIR(3.74±3.59% vs 4.95±3.48%,p=0.034),相关性更好(r=0.60 vs r=0.47),并显著降低图像噪声和线束硬化伪影。

要点

  • MBIR较HIR显著降低图像噪声(3.4±0.7 vs 2.1±0.8,p<0.001)和线束硬化伪影(3.8±0.4 vs 2.6±1.0,p<0.001)。
  • MBIR的CT与MRI ECV平均绝对差显著小于HIR(3.74±3.59% vs 4.95±3.48%,p=0.034)。
  • MBIR的CT与MRI ECV相关性优于HIR(r=0.60,p<0.001 vs r=0.47,p=0.006)。

核心结论

  • 含线束硬化模型的MBIR较HIR有效降低图像噪声和线束硬化伪影,以MRI为参考标准时改善心肌ECV定量。

研究设计与样本

  • 回顾性研究,共纳入34例患者,均行心脏CT和MRI检查。
  • 利用HIR和含线束硬化模型的MBIR分别创建配对CT图像集。
  • 以MRI为参考标准,通过Pearson相关分析计算CT与MRI全局中间心室ECV的平均绝对差和相关性。
  • 对增强后图像进行图像噪声和线束硬化伪影的定性分析,采用四点量表:1=广泛,2=强,3=轻度,4=极小。

关键影像表现/征象

  • 定性分析显示MBIR较HIR显著降低图像噪声(MBIR 3.4±0.7 vs HIR 2.1±0.8,p<0.001)。
  • MBIR较HIR显著降低线束硬化伪影(MBIR 3.8±0.4 vs HIR 2.6±1.0,p<0.001)。

诊断标准与关键数值

  • CT与MRI全局中间心室ECV平均绝对差:MBIR 3.74±3.59% vs HIR 4.95±3.48%,p=0.034。
  • CT与MRI ECV相关性:MBIR r=0.60(p<0.001)vs HIR r=0.47(p=0.006)。
  • 图像噪声和线束硬化伪影定性评分采用四点量表:1=广泛,2=强,3=轻度,4=极小。

临床应用与阅片要点

  • 含线束硬化模型的MBIR较HIR可降低图像噪声和线束硬化伪影,并改善心肌ECV定量。
  • 以MRI为参考标准时,MBIR的CT与MRI ECV平均绝对差小于HIR,相关性优于HIR。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare the effects of hybrid iterative reconstruction (HIR) and model-based iterative reconstruction (MBIR) that incorporates a beam-hardening model for myocardial extracellular volume (ECV) quantification by cardiac CT using MRI as a reference standard.

METHODS: In this retrospective study, a total of 34 patients were evaluated using cardiac CT and MRI. Paired CT image sets were created using HIR and MBIR with a beam-hardening model. We calculated mean absolute differences and correlations between the global mid-ventricular ECV derived from CT and MRI via Pearson correlation analysis. In addition, we performed qualitative analysis of image noise and beam-hardening artifacts on postcontrast images using a four-point scale: 1 = extensive, 2 = strong, 3 = mild, and 4 = minimal.

RESULTS: The mean absolute difference between the ECV derived from CT and MRI for MBIR was significantly smaller than that for HIR (MBIR 3.74 ± 3.59%; HIR 4.95 ± 3.48%, p = 0.034). MBIR improved the correlation between the ECV derived from CT and MRI when compared with HIR (MBIR, r = 0.60, p < 0.001; HIR, r = 0.47, p = 0.006). In qualitative analysis, MBIR significantly reduced image noise and beam-hardening artifacts when compared with HIR ([image noise, MBIR 3.4 ± 0.7; HIR 2.1 ± 0.8, p < 0.001], [beam-hardening artifacts, MBIR 3.8 ± 0.4; HIR 2.6 ± 1.0, p < 0.001]).

CONCLUSIONS: MBIR with a beam-hardening model effectively reduced image noise and beam-hardening artifacts and improved myocardial ECV quantification when compared with HIR using MRI as a reference standard.

KEY POINTS: • MBIR with a beam-hardening model effectively reduced image noise and beam-hardening artifacts. • The mean absolute difference between the global mid-ventricular ECV derived from CT and MRI for MBIR was significantly smaller than that for conventional HIR. • MBIR provided more accurate myocardial CT number and improved ECV quantification when compared with HIR.

来源:Eur Radiol,查看原文。版权归原作者所有。