文献速递胸部技术与安全DR

利用计算机模拟研究数字化X射线摄影胸部成像中平均体型患者的优化管电压

Use of a computer simulator to investigate optimized tube voltage for chest imaging of average patients with a digital radiography (DR) imaging system.
PubMed 文献 · Br J Radiol · 2019-10-07

本研究通过计算机模拟生成含临床相关解剖结构的数字化X射线摄影胸部影像,由5名经验丰富的图像评估者盲法对20例患者在不同管电压下的整体质量、肋骨、肺、肺门、脊柱、膈肌及肺结节标准进行评分,并与参考管电压图像比较。结果显示整体质量、肺门、膈肌和脊柱标准在低及高管电压下表现差,峰值在80-100 kVp;肋骨标准在低kVp下更优;肺和肺结节标准变化小。结论提示80-100 kVp的中段管电压对平均成人最优。

要点

  • 整体质量、肺门、膈肌和脊柱标准在80-100 kVp达峰值,低或高管电压下评分下降。
  • 肋骨标准在低kVp下表现更优,肺和肺结节标准随管电压变化小。
  • 模拟研究推荐平均成人胸部DR的优化管电压范围为80-100 kVp。

核心结论

  • 虚拟临床试验结果表明,80-100 kVp的中段管电压对平均成人胸部DR成像最优。

研究设计与样本

  • 计算机模拟研究,采用数字重建放射影像算法生成含临床相关解剖的胸部DR图像;5名经验丰富的图像评估者对20例患者在不同管电压下的图像按临床图像标准评分,并与同一患者参考kVp重建图像比较,评估者盲法;采用线性混合效应模型分析。

关键影像表现/征象

  • 整体质量、肺门、膈肌和脊柱标准在低及高管电压下表现差,峰值在80-100 kVp。
  • 肋骨标准在低kVp下表现更优。
  • 肺和肺结节标准随管电压变化小。

诊断标准与关键数值

  • 评估标准包括整体质量、肋骨、肺、肺门、脊柱、膈肌和肺结节,均与同一患者参考kVp重建图像比较评分。
  • 整体质量、肺门、膈肌和脊柱标准在80-100 kVp达峰值。

临床应用与阅片要点

  • 研究结果提示80-100 kVp的中段管电压对平均成人胸部DR成像最优。
  • 原文指出目前尚无胸部DR成像优化管电压参数的具体推荐。

局限性

  • 原文指出目前尚无胸部DR成像优化管电压参数的具体推荐。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The aim of this study was to investigate via computer simulation a proposed improvement to clinical practice by deriving an optimized tube voltage (kVp) range for digital radiography (DR) chest imaging.

METHODS: A digitally reconstructed radiograph algorithm was used which was capable of simulating DR chest radiographs containing clinically relevant anatomy. Five experienced image evaluators graded clinical image criteria, i.e. overall quality, rib, lung, hilar, spine, diaphragm and lung nodule in images of 20 patients at tube voltages across the diagnostic energy range. These criteria were scored against corresponding images of the same patient reconstructed at a specific reference kVp. Evaluators were blinded to kVp. Evaluator score for each criterion was modelled with a linear mixed effects algorithm and compared with the score for the reference image.

RESULTS: Score was dependent on tube voltage and image criteria in a statistically significant manner for both. Overall quality, hilar, diaphragm and spine criteria performed poorly at low and high tube voltages, peaking at 80-100 kVp. Lung and lung nodule demonstrated little variation. Rib demonstrated superiority at low kVp.

CONCLUSION: A virtual clinical trial has been performed with simulated chest DR images. Results indicate mid-range tube voltages of 80-100 kVp are optimum for average adults.

ADVANCES IN KNOWLEDGE: There are currently no specific recommendations for optimized tube voltage parameters for DR chest imaging. This study, validated with images containing realistic anatomical noise, has investigated and recommended an optimal tube voltage range.

来源:Br J Radiol,查看原文。版权归原作者所有。